The preset puncture frequency amounts to 100 punctures per second. The controller auto-
matically resets to this preset every time it is switched on. The device has a puncture fre-
quency range of 50-150 punctures per second and can be adjusted in 10 steps.
The number of penetrations per second shown on the display may deviate by ±10 % in sta-
tionary operation from the value that actually occurs at the handpiece.
To set a suitable penetration frequency:
E
Ensure that the controller is switched on (see chapter "6.1 Switching the control unit to
operating mode or standby mode" on page 27).
E
Press the
Increase frequency
(
) button to increase the penetration frequency.
E
Press the
Decrease frequency
(
) button to decrease the penetration frequency.
Switching the handpiece on and off
CAUTION
Danger of injury when using the handpiece
If an unsecured handpiece is switched on, it may move uncontrollably through vibrations
and cause injury.
E
Hold the handpiece firmly before switching it on or place it in the handpiece tray.
To switch on the handpiece:
E
Ensure that the power supply is established (see the chapter "5.2 Connecting the power
supply" on page 21).
E
Ensure that the controller is switched on (see chapter "6.1 Switching the control unit to
operating mode or standby mode" on page 27).
E
Ensure that the handpiece is correctly connected (see chapter "5.4 Connecting the
handpiece" on page 22).
E
Select the handpiece connected with the help of the
Select handpiece
(
) but-
tons.
E
Press the
GO
button (
) on the controller.
The LED indicator next to the button illuminates white and shows the current operation
of the handpiece.
If the handpiece does not start:
E
Refer to the chapter "10 Questions and problems" on page 33.
To switch off the handpiece:
E
Press the
GO
(
) button again.