Chapter 6: Maintaining your XPert or XVS Enclosure
Product Service 1-800-522-7658
36
Digital 1000 Calibration
1. Calibrate the airflow monitor according to the instruction manual that
comes with the kit. To successfully calibrate, it will be necessary to
change the face velocity by adjusting the airflow exhaust volume.
The face velocity can also be changed by raising and lowering the
sash. The exhaust volume can be adjusted with the speed control on
the FilterMate or by using an adjustable damper on the exhaust
blower. Typical calibration conditions are set at face velocity air
sample differences of at least 20 feet per minute. The airflow monitor
is factory set to be calibrated with a difference of at least 50 fpm and
can be changed by changing the “lower/higher air sample difference”.
The following suggested in flow face velocity speeds are
recommended to successfully calibrate. Typical low air alarms are set
10-20 fpm below operational speeds. Follow Step 2 below and
review the Labconco 1000 Alarm User’s Manual that comes with the
airflow monitor.
Low Air
Alarm
Set Point
Enclosure
Operating In
flow Speed
Low
Calibration
Set Point
High
Calibration
Set Point
40 - 50 fpm
60 fpm
40 - 60 fpm
100 – 120 fpm
60 - 70 fpm
80 fpm
50 - 90 fpm
100 – 150 fpm
80 – 90 fpm
100 fpm
50 – 110 fpm
100 – 170 fpm
2. Go to SETUP menu and then CAL CONFIG MENU and change the
“lower/higher air sample difference” to 20 fpm. This will allow you
to successfully calibrate with values of a minimum of 20 fpm
difference.
3. While in CAL CONFIG MENU, change the “sensor difference” from
10% to 3%.
4. While in CAL CONFIG MENU, adjust the red low air alarm to the
desired setting such as 60 fpm. Then adjust the yellow “CAUTION
or WARNING” to 63 fpm, if desired. Then adjust the “CAUTION or
WARNING” air reset to 3 fpm. This sets the alarm condition.
5. To complete the CAL CONFIGURATION, be sure to enter “DONE”.
6. To start the calibration mode, use the Labconco 1000 Manual and
enter “CALIBRATION” mode on the display from the SETUP menu.
7. Follow the instructions on the display and alter the low exhaust
volume with the speed control on the FilterMate or exhaust damper;
alternatively, the face velocity may be changed by raising the sash.
Measure the average face velocity and enter the low value on the
display. Be careful not to block the opening. The low exhaust
volume calibration will take about 5 seconds.
Summary of Contents for XPert Series
Page 49: ...Chapter 7 Accessorizing your XPert or XVS Enclosure Product Service 1 800 522 7658 45...
Page 54: ...Appendix A Replacement Parts Product Service 1 800 522 7658 50...
Page 56: ...Appendix B Dimensions Product Service 1 800 522 7658 52 2 3 and 4 Units...
Page 57: ...Appendix B Dimensions Product Service 1 800 522 7658 53 2 3 and 4 Units...
Page 58: ...Appendix B Dimensions Product Service 1 800 522 7658 54...