Chapter 3: Getting Started
Product Service 1-800-522-7658
13
Connecting to the Vented Enclosure Exhaust
System
WARNING: The weight of the exhaust ductwork system must be
supported independently of the enclosure superstructure or damage may
occur.
The exhaust system should be installed by a qualified HVAC contractor.
The exhaust connection on the 2' through 6' enclosures has been designed to
accept a 2" x 10" (5.1cm x 25.4cm) nominal transition adapter. The 8'
enclosure has two 2" x 10" connections. See Chapter 7 for ordering
accessory transition adapters. Labconco manufactures transition adapters for
either top or bottom exhaust and for both 5" dia. hose or 6" dia. duct. Review
Chapter 2 for remote blower and FilterMate Portable Exhauster exhaust
prerequisites and review Chapter 7 for ordering blower exhaust equipment.
For your convenience several exhaust options are shown in Figures 3-2, 3-3,
3-4 and 3-5. Consult Labconco Customer Service should you require help
sizing your blower for the exhaust volume and system static pressure loss.
To ensure compatibility, the selected exhaust duct material should match the
enclosures, procedures and chemical applications.
Figure 3-2 Top Connection to Outside Exhaust
6.03" ID Exhaust
Transition Adapter
Summary of Contents for XPert Series
Page 49: ...Chapter 7 Accessorizing your XPert or XVS Enclosure Product Service 1 800 522 7658 45...
Page 54: ...Appendix A Replacement Parts Product Service 1 800 522 7658 50...
Page 56: ...Appendix B Dimensions Product Service 1 800 522 7658 52 2 3 and 4 Units...
Page 57: ...Appendix B Dimensions Product Service 1 800 522 7658 53 2 3 and 4 Units...
Page 58: ...Appendix B Dimensions Product Service 1 800 522 7658 54...