![Labconco XPert Series User Manual Download Page 24](http://html1.mh-extra.com/html/labconco/xpert-series/xpert-series_user-manual_3306498024.webp)
Chapter 4: Performance Features and Safety Precautions
Product Service 1-800-522-7658
20
1.
Aerodynamic Clean-Sweep™ Air Foil
has a unique low profile
shape that allows air to sweep the work surface for maximum
containment. The Clean-Sweep™ openings create a constant
protective barrier from contaminants. In addition, should the operator
inadvertently block the airflow entering the airfoil, air continues to
pass under the air foil and through the Clean-Sweep openings. See
Figure 4-2.
2.
Containment-Enhancing Upper Sash Foil
includes a perforated air
passage directly atop the sash foil to bleed air into the hood chamber
and direct chemical and powder concentrations away from the sash
opening. The radiused sash foil sweeps airflow into the hood with
minimal turbulence. See Figure 4-3.
Figure 4-2
Figure 4-3
Summary of Contents for XPert Series
Page 49: ...Chapter 7 Accessorizing your XPert or XVS Enclosure Product Service 1 800 522 7658 45...
Page 54: ...Appendix A Replacement Parts Product Service 1 800 522 7658 50...
Page 56: ...Appendix B Dimensions Product Service 1 800 522 7658 52 2 3 and 4 Units...
Page 57: ...Appendix B Dimensions Product Service 1 800 522 7658 53 2 3 and 4 Units...
Page 58: ...Appendix B Dimensions Product Service 1 800 522 7658 54...