Chapter 3: Getting Started
Product Service 1-800-522-7658
13
Original instructions
Connecting the Electrical Supply Source
to the Fiberglass 30 Fume Hood
Prior to connecting any electrical wiring to the fume hood structure, refer to the
hood identification plate for the proper electrical requirements of your specific
model.
WARNING
:
The building electrical supply system for
Fiberglass 30 Hoods is required to include overload protection.
A switch or circuit breaker shall be in close proximity to the
equipment and within easy reach of the operator. The switch or
circuit breaker is to be marked as the disconnecting device for
the equipment. Do not position the unit so that it is difficult to
operate the main disconnect device.
AVERTISSEMENT :
Le système d'alimentation électrique
pour la construction en fibre de verre 30 Hoods est nécessaire
d'inclure la protection de surcharge
. Un commutateur ou
disjoncteur doit être tout près de l'équipement et à portée facile
de l'opérateur. Le commutateur ou le disjoncteur doit être
marqué comme l'appareil débranchant pour l'équipement.
Ne
pas placer l'appareil de sorte qu'il est difficile de faire
fonctionner le dispositif de déconnexion principal.
The identification plate, model number, serial number, and electrical connection
boxes are accessible from the front of the fume hood by removing the front panel.
The Fiberglass 30 Hood is normally wired for 115 Volt, 60 Hz, 20 Amp or 230
Volt, 50 Hz, 10 Amp electrical services. The number of circuits varies depending
on the model. All of the electrical connections are terminated at the single point
internal junction box for hook-up by a qualified electrician. Refer to the wiring
diagram for your Fiberglass 30 Hood in
Appendix C: Fiberglass 30 Fume Hood
Specifications.
!