63
Wing Work Surface Removal
Before beginning, the center work surface must be hinged open or removed as
described previously. Reference Figures 7-1 and 8-2 for part identification and follow
these steps.
1. If not already on, turn the cabinet blower on.
2. Lift either wing work surface approximately 2 inches (50 mm) by grasping the
knobs at either front corner.
3. Grasp the front edge of the wing work surface and pull it forward to free the
locating tab at the rear edge from the back wall of the cabinet.
If removing either wing work surface from the cabinet, the wing work surface must be
thoroughly decontaminated before removing it from the cabinet.
Once the wing work surfaces are removed, the center drip tray will remain, see Figure
8-3.
To reinstall the wing work surface, engage the locating tab on the rear edge with the slot
in the back wall of the cabinet. Gently hinge the wing down. Take care to engage the
pin in the front corner of the wing work surface with the hole in the grille.
Center Drip Tray
Knob
Front Grille
Exhaust Duct Flange
Alignment Tab
Figure 8-3