The exhaust system must be capable of moving the following volumes of exhaust air at
the negative pressures listed. The
Airflow Volumes
are the values recorded via direct
measurement using a flow hood at the front opening of the cabinet. The
Concurrent
Balance Values
are measured in the exhaust duct via traverse methodology, and will
always be higher due to differences in volume measurement methodologies.
1:
Unlike Type B biosafety cabinets, the recommended vacuum will remain constant throughout the life of
the exhaust HEPA filter. Duct vacuums below 0.2 inches H
2
O (50 Pa) or above 0.5 inches H
2
O (125 Pa)
may result in erratic operation and throw an alarm condition from the product.
2:
WC = Inches of Water Column, typically expressed in units of
inches H
2
O
.
Air-Tight Damper
Axiom Type C1 Model
Airflow Volume
Concurrent
Balance Value
Recommended
Duct Vacuum
1
ft
3
/min
m
3
/hr
ft
3
/min
m
3
/hr
WC
2
Pa
4-foot, 8" Sash
323
549
387
658
0.30
75
4-foot, 10" Sash
400
680
480
816
0.30
75
6-foot, 8" Sash
463
787
556
945
0.30
75
6-foot, 10" Sash
570
968
684
1162
0.30
75
Values are Nominal Inflow (105 fpm) + 50 cfm through Canopy Inlet Relief Valve. CBV = ~ 20% higher
Table 2-1
AIR-TIGHT DAMPER
DUCT COLLAR
BUILDING EXHAUST DUCT
(NOT INCLUDED)
AIRFLOW
ADJUSTMENT
HANDLE
BIOSAFETY CABINET
DUCT STUB
Figure 2-6
Visual appearance of biosafety cabinet and exhaust connection may vary by model.
Summary of Contents for Axiom C1
Page 2: ...Axiom Type C1 Biosafety Cabinets 2021 Present 30441xxx1 30448xxx1 30461xxx1 30468xxx1...
Page 75: ...75 Blower only will not start...
Page 76: ...76 Lights only will not illuminate...
Page 77: ...77 UV Light will not illuminate...
Page 78: ...78 Airflow Alert activating...
Page 79: ...79 Filter Life Gauge not at 100 when new...
Page 80: ...80 Contamination in the work area...
Page 82: ...82 Figure A 1 1 2 3 4...