45
UV Parameters
For models equipped with the optional UV light, the cabinet has an integral UV light
maintenance system. It allows you to monitor how many hours the lamp has been on, to
reset the UV lamp hourmeter, and to define how many hours you want the UV lamp to
operate before receiving a reminder to replace it.
This screen displays the hours of UV lamp operation
(Runtime), and how many hours remain (Remaining)
until you receive a warning to replace the lamp. To
reset the Runtime hourmeter to zero (after replacing
the UV lamp), select
Reset
, then
[OK/MUTE]
. The
hour text will begin to flash, if you entered this
condition by mistake, press
[MENU]
. If you want to
reset the hourmeter, hold
[OK/MUTE]
for 3 seconds.
To change the desired UV lamp lifetime, select
Lifetime
as seen in Fig. 6-20, then
[OK/MUTE]
. The
screen shown in Fig. 6-21 will display. To change the
UV lamp lifetime (number of operating hours before
receiving a warning), change the
Hour
field accordingly
using
[UP]
or
[DOWN]
, then
[OK/MUTE]
.
For most UV lamps, the output of UV light decreases at a constant rate. Typically, after
6,000 hours of operation the lamp
’s output intensity will reduce to 80% of when it was
new. This option allows you to set operational life of the UV lamp, in 100 hour
increments. 6,000 hours is the default.
The Tools Submenu
This submenu is reserved for use by certifiers, during certification or service
procedures.
CAUTION!
- Entering this submenu will disable some alarms and
functionality so that diagnostic and certification procedures can be performed.
Additional details on the Tools Submenu are found in the Axiom Type C1 Technical
Manual.
RUNTIME
2139 H
REMAINING
3861 H
6000 H
SET LIFETIME
Reset
Lifetime
Figure 6-20
Figure 6-21
Summary of Contents for Axiom C1
Page 2: ...Axiom Type C1 Biosafety Cabinets 2021 Present 30441xxx1 30448xxx1 30461xxx1 30468xxx1...
Page 75: ...75 Blower only will not start...
Page 76: ...76 Lights only will not illuminate...
Page 77: ...77 UV Light will not illuminate...
Page 78: ...78 Airflow Alert activating...
Page 79: ...79 Filter Life Gauge not at 100 when new...
Page 80: ...80 Contamination in the work area...
Page 82: ...82 Figure A 1 1 2 3 4...