Chapter 8: Troubleshooting
PROBLEM
CAUSE
CORRECTIVE ACTION
Airflow Monitor
Malfunction
(Continued)
Wrong alarm set
point.
Airflow monitor was not properly
adjusted. Repeat calibration steps
outlined in this manual in Chapter 6.
Constant
audible
alarm.
Check airflow and calibration of
airflow monitor. See Chapter 6.
Continuous alarm.
Check the downflow velocity of the
enclosure as the airflow of the
system may have changed. If
incorrect, adjust the speed control to
increase downflow velocity. The
HEPA filter may have become
loaded. If downflow velocity is
correct, calibrate the airflow monitor
outlined in this manual in Chapter 6.
Monitor alarms; air
way to airflow
monitor sensor is
blocked by insects,
dust or debris.
Lightly clean the airway with clean
air. Be careful not to touch sensitive
electrical components.
Audible
disable
will not stay
operational.
An alarm condition must be
continuously present before the
audible alarm can be silenced. If
flow conditions fluctuate near the
alarm set point, the airflow monitor
will automatically reset it. Action
should be taken to bring the
enclosure airflow into proper
operating parameters or adjust the
alarm set point lower.
Summary of Contents for 3970202
Page 2: ......
Page 23: ...18 Chapter 4 Performance Features and Safety Precautions Figure 4 3 UV Sash Closure 16...
Page 58: ...53 Appendix B Dimensions and Exhaust Options Figure B 1 Purifier Filtered PCR Enclosure...
Page 60: ...55 Appendix C Filtered Enclosure Specifications C 1...
Page 61: ...56 Appendix C Filtered Enclosure Specifications C 2...
Page 62: ...57 Appendix C Filtered Enclosure Specifications C 3...
Page 63: ...58 Appendix C Filtered Enclosure Specifications C 4...
Page 65: ...60...
Page 66: ......