19
Chapter 4: Performance Features and Safety Precautions
1.
Aerodynamic Lower Air Foil
allows air to sweep the
work surface and promote ISO Class 5 air conditions inside
the enclosure. See Figure 4-1.
2.
Ergonomic Slope
of 10 degrees provides maximum
visibility and comfort, reduces glare, thereby minimizing
operator fatigue. See Figure 4-1.
3.
Internal Depth of 25"
provides necessary depth to work
inside the enclosure. See Figure 4-1.
4.
Safety Glass Sash with Spring-Loaded Latch
has a
wiping seal and features a spring-loaded latch to secure
sash open for loading and cleaning. The sash must be
down for normal operation. See Figure 4-1.
5.
Utility Ports with Iris
allows electrical cords and data
cords to pass through the back of enclosure without leaving
a large hole. The enclosure ships with solid plugs. Iris
plugs are included with the User’s Manual. See Figure 4-1
and 4-2.
6.
Optional Guardian™ Airflow Monitor
continuously
monitors airflow. An audio/visual alarm alerts the user to
low airflow conditions. The Guardian™ Airflow Monitor
is an available option on all Purifier Filtered PCR
Enclosure models. See Figure 4-1 and 4-2.
7.
Inherently Safe Impeller
has a negative pressure plenum
that surrounds the positive pressure impeller so that if a
leak should occur, the unfiltered air is captured and
refiltered. See Figure 4-2.
8.
Speed Control
regulates the speed of the motorized
impeller and is used by the certifier to validate and adjust
the downflow velocity. Located behind the front panel.
See Figures 4-1 and 4-2.
9.
Fluorescent Lamp
is located above the work area, out of
contact with the clean air. A safety glass window beneath
the lamp distributes the light evenly across the work
surface. See Figure 4-1.
10.
Filter Pressure Gauge
indicates the pressure across the
HEPA filter to help predict HEPA filter loading. See
Figure 4-1.
Summary of Contents for 3970202
Page 2: ......
Page 23: ...18 Chapter 4 Performance Features and Safety Precautions Figure 4 3 UV Sash Closure 16...
Page 58: ...53 Appendix B Dimensions and Exhaust Options Figure B 1 Purifier Filtered PCR Enclosure...
Page 60: ...55 Appendix C Filtered Enclosure Specifications C 1...
Page 61: ...56 Appendix C Filtered Enclosure Specifications C 2...
Page 62: ...57 Appendix C Filtered Enclosure Specifications C 3...
Page 63: ...58 Appendix C Filtered Enclosure Specifications C 4...
Page 65: ...60...
Page 66: ......