LabCo 400.003.050 User Manual Download Page 2

 

 

CONTENTS 

PAGE NUMBER 

INTRODUCTION 

WARRANTY 

DELIVERY 

SAFETY REMINDER 

1. SPECIFICATIONS 

2. DECLARATION OF CONFORMITY 

3. REQUIRED OPERATIONAL CONDITIONS 

4. INSTALLATION 

5. STRUCTURE 

6. OPERATION PANEL 

7. ROTOR PREPARATION 

8. OPERATION 

10 

9. MAINTENANCE 

14 

10. TROUBLESHOOTING 

15 

11. INSTRUCTIONS OF ROTOR AND TUBE 

16 

12. CALCULATE RCF 

18 

AFTER-SALES SERVICE 

18 

 
 

 
 

 

 

 

 

Summary of Contents for 400.003.050

Page 1: ...MINI CENTRIFUGE USER MANUAL sales labcoscientific com au 1800 052 226 ABN 57 622 896 593 labcoscientific com au PO Box 5816 Brendale QLD 4500 Version 1 1...

Page 2: ...FICATIONS 5 2 DECLARATION OF CONFORMITY 5 3 REQUIRED OPERATIONAL CONDITIONS 6 4 INSTALLATION 6 5 STRUCTURE 7 6 OPERATION PANEL 8 7 ROTOR PREPARATION 9 8 OPERATION 10 9 MAINTENANCE 14 10 TROUBLESHOOTIN...

Page 3: ...as been operated and maintained properly Warranty of the Rotor Please pay attention do not use the rotor once it has been corrosion or fatigue damage We do not guarantee this centrifuge and the rotor...

Page 4: ...centrifuge in any manner not described in this User manual When in doubt or have any troubles with this centrifuge ASK FOR HELP The precautions described in this User manual are carefully developed i...

Page 5: ...ices to the centrifuge except by our service centre are strictly prohibited CAUTION This centrifuge must be located on one firm and level table Make sure the centrifuge is horizontal before running Ma...

Page 6: ...omplies with Canadian ICES 001 SPECIFICATIONS DESCRIPTION MAXIMUM SPEED 15000rpm increment 100rpm MAXIMUM RCF 15100 g increment 100 g MAXIMUM CAPACITY 2ml 12 TIMER 30 seconds 99 minutes HOLD continuou...

Page 7: ...ge the rotor must be removed Location 1 Locate this centrifuge on a firm flat and level table ensure the four feet of this centrifuge stand on the table firmly Avoid installing on a slippery table boa...

Page 8: ...utlet 2 A cover 10A outlet which has a good ground is required and meets with local safety requirements 5 STRUCTURE Figure 5 1 400 003 050 front view of centrifuge Door lock hook Rotor Operation panel...

Page 9: ...8 Figure 5 2 D2012 Rear view of centrifuge 6 OPERATION PANEL Speed area Lock status Time area 150 x 100rpm 12 min 1 2 3 4 5 Figure 6 1 Operation Panel Air vents View Door Power inlet Power switch...

Page 10: ...utton Press down the button to start running The centrifuge will stop running if the button is pressed during centrifugation 1 Speed area Lock status Time area 150 x 100rpm 12 min Figure 6 2 the main...

Page 11: ...m 12 min Figure 8 1 the last running interface o Speed 15000rpm Running time 12 minutes o The door lock is released CAUTION Any abnormalities such as corrosion or scratches are found stop using the ro...

Page 12: ...firmly Put the rotor lid on the rotor press the fastener down and make the fastener join with the rotor nut tightly Close the door and start running To release the rotor first pull the fastener up to...

Page 13: ...m minimum counter clockwise rotate parameter button values circulate from large small minimum maximum 2 Set the time Press select button time value flash into time setting mode Rotate parameter button...

Page 14: ...ol will flash into RCF parameter input status If no button is pressed after the speed parameter has flashed for 5 seconds the inputting mode will be shut down 1 Rotate parameter button to input a RCF...

Page 15: ...can wipe the drive shaft with soft cloth and then apply a thin coat of silicon grease to it 4 Door Clean and sterilise the door using the same method as the step 1 5 Rotor To prevent corrosion remove...

Page 16: ...e door opening Close the door immediately Close the door and then start to operate E 06 Set wrong speed The setting speed exceeds the allowable range Modify the speed value E 10 86 Read the service ma...

Page 17: ...on the top right side of the unit Insert a door key into the hole and push forward to release the door lock and then open the door 11 INSTRUCTIONS OF ROTOR AND TUBE The rotor instructions Rotor struct...

Page 18: ...can be high pressure sterilization 121 1 0kg cm2 20 minutes Cleaning and sterilising tubes please refer to the table below PA Polyallomer PC Polycarbonate PP Polypropylene O Applicable X Inapplicable...

Page 19: ...ure before the tubes are removed 11 2 4 CONDITION AND LIFE EXPECTANCY OF TUBES The life expectancy of plastic tubes depends on the characteristics of samples speed of the rotor used and temperature ap...

Page 20: ...19...

Reviews: