
14
9. MAINTENANCE
1. Centrifuge
•
If the centrifuge is exposed to ultraviolet rays for a long time, the color of the doors may change and/or the label
may come off. After using, cover the centrifuge with a piece of cloth to protect it from direct exposure.
•
Once the centrifuge is heavily dirty, clean it with a cloth or sponge moistened with a neutral detergent solution.
•
Sterilise the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
2. Rotor chamber
•
Once the rotor chamber is dirty, clean with cloth or sponge moistened with a neutral detergent solution. Sterilise
the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
3. Drive shaft
•
It’s recommended having regular maintenance for the drive shaft. You can wipe the drive shaft with soft cloth, and
then apply a thin coat of silicon grease to it.
4. Door
•
Clean and sterilise the door using the same method as the step 1).
5. Rotor
•
To prevent corrosion, remove the rotor from rotor chamber. If do not use regularly then detach the rotor door and
flip the rotor to dry the tube holes.
•
If a sample is leaked in the rotor, rinse the rotor with water. Apply a thin coat of silicon grease to the rotor when it
is completely dried.
•
The rotor should have regular maintenance, its recommended cleaning it each 3 months to ensure tube and rotor
holes keep clean, and then apply a thin coat of silicon grease.
Not following the recommended instructions for cleaning or disinfecting may damage the centrifuge.
CAUTION
:
Do not directly pour water, neutral detergent or disinfectant solution into the rotor chamber. Otherwise
fluids may leak into the drive units and cause corrosion or deterioration to the bearings.
Summary of Contents for 400.003.050
Page 20: ...19...