10. TROUBLESHOOTING 

Frequent Problems List 

This centrifuge is designed with self-diagnosing function. When something is wrong and it is not available to operate, the 
alarm code will be displayed on the screen, and the operator can know the malfunction caused by the alarm code. 

 

SYMPTOM 

CAUSES 

SOLUTIONS 

NOTHING APPEARS ON THE SCREEN 

WHEN THE POWER IS TURNED ON. 

·Building power circuit breaker trips. 
·the fuse was blown out. 

-Replace the fuse. 

ABNORMAL VIBRATION 

·Rotor do not match with spindle 
·Samples are imbalance 
·Rotor lid loose 

·Install the rotor again  
·Weighting scales, install 

symmetrically  
·Tighten the rotor lid firmly 

 AL
AR
M

 C

O

DE
 AP
P

E

AR
E

ON
 T

H

E

 T

IM

E

 D

IS

P

LA

Y

    

 

SC
R

EEN

 

E-02 
Door fault

 

·The door opened in running. 

·The 

button is pressed while 

the door opening. 

·Close the door immediately. 
·Close the door

and then start to 

operate. 

E-06 
Set wrong speed 

 

·The setting speed exceeds the 

allowable range. 

·Modify the speed value. 

E-10

86

 

·Read the service manual 

·Contact with service centre 

Table 10-1 frequent problems and solutions 

 

 

Alarm  code  E-1

9  are  related  to  wrong  operating.  You  can  continue  running  the  centrifuge  after  the  cause 

removed.

 

 

How to open the door 

 

In the case of

 

power on 

 

 

Turn on the POWER switch, the door lock will release automatically. 

 

The door lock will release automatically once the operation finished. 

 

While the rotor stops, press

button, door lock can be released to open the door. 

 

 WARNING: Do not open the door when the rotor is rotating during maintenance. 

 WARNING: 

 

Never open the door while the centrifuge is running. 

 

The door is opened once rotor is still rotating, close the door immediately. 

 CAUTION: The door can be opened while the power is on and the rotor stops rotating 

Summary of Contents for 400.003.050

Page 1: ...MINI CENTRIFUGE USER MANUAL sales labcoscientific com au 1800 052 226 ABN 57 622 896 593 labcoscientific com au PO Box 5816 Brendale QLD 4500 Version 1 1...

Page 2: ...FICATIONS 5 2 DECLARATION OF CONFORMITY 5 3 REQUIRED OPERATIONAL CONDITIONS 6 4 INSTALLATION 6 5 STRUCTURE 7 6 OPERATION PANEL 8 7 ROTOR PREPARATION 9 8 OPERATION 10 9 MAINTENANCE 14 10 TROUBLESHOOTIN...

Page 3: ...as been operated and maintained properly Warranty of the Rotor Please pay attention do not use the rotor once it has been corrosion or fatigue damage We do not guarantee this centrifuge and the rotor...

Page 4: ...centrifuge in any manner not described in this User manual When in doubt or have any troubles with this centrifuge ASK FOR HELP The precautions described in this User manual are carefully developed i...

Page 5: ...ices to the centrifuge except by our service centre are strictly prohibited CAUTION This centrifuge must be located on one firm and level table Make sure the centrifuge is horizontal before running Ma...

Page 6: ...omplies with Canadian ICES 001 SPECIFICATIONS DESCRIPTION MAXIMUM SPEED 15000rpm increment 100rpm MAXIMUM RCF 15100 g increment 100 g MAXIMUM CAPACITY 2ml 12 TIMER 30 seconds 99 minutes HOLD continuou...

Page 7: ...ge the rotor must be removed Location 1 Locate this centrifuge on a firm flat and level table ensure the four feet of this centrifuge stand on the table firmly Avoid installing on a slippery table boa...

Page 8: ...utlet 2 A cover 10A outlet which has a good ground is required and meets with local safety requirements 5 STRUCTURE Figure 5 1 400 003 050 front view of centrifuge Door lock hook Rotor Operation panel...

Page 9: ...8 Figure 5 2 D2012 Rear view of centrifuge 6 OPERATION PANEL Speed area Lock status Time area 150 x 100rpm 12 min 1 2 3 4 5 Figure 6 1 Operation Panel Air vents View Door Power inlet Power switch...

Page 10: ...utton Press down the button to start running The centrifuge will stop running if the button is pressed during centrifugation 1 Speed area Lock status Time area 150 x 100rpm 12 min Figure 6 2 the main...

Page 11: ...m 12 min Figure 8 1 the last running interface o Speed 15000rpm Running time 12 minutes o The door lock is released CAUTION Any abnormalities such as corrosion or scratches are found stop using the ro...

Page 12: ...firmly Put the rotor lid on the rotor press the fastener down and make the fastener join with the rotor nut tightly Close the door and start running To release the rotor first pull the fastener up to...

Page 13: ...m minimum counter clockwise rotate parameter button values circulate from large small minimum maximum 2 Set the time Press select button time value flash into time setting mode Rotate parameter button...

Page 14: ...ol will flash into RCF parameter input status If no button is pressed after the speed parameter has flashed for 5 seconds the inputting mode will be shut down 1 Rotate parameter button to input a RCF...

Page 15: ...can wipe the drive shaft with soft cloth and then apply a thin coat of silicon grease to it 4 Door Clean and sterilise the door using the same method as the step 1 5 Rotor To prevent corrosion remove...

Page 16: ...e door opening Close the door immediately Close the door and then start to operate E 06 Set wrong speed The setting speed exceeds the allowable range Modify the speed value E 10 86 Read the service ma...

Page 17: ...on the top right side of the unit Insert a door key into the hole and push forward to release the door lock and then open the door 11 INSTRUCTIONS OF ROTOR AND TUBE The rotor instructions Rotor struct...

Page 18: ...can be high pressure sterilization 121 1 0kg cm2 20 minutes Cleaning and sterilising tubes please refer to the table below PA Polyallomer PC Polycarbonate PP Polypropylene O Applicable X Inapplicable...

Page 19: ...ure before the tubes are removed 11 2 4 CONDITION AND LIFE EXPECTANCY OF TUBES The life expectancy of plastic tubes depends on the characteristics of samples speed of the rotor used and temperature ap...

Page 20: ...19...

Reviews: