background image

Base Station 
With two foldable legs at the back of the unit, the base station can be 
placed onto any flat surface or wall mounted at the desired location by 
the hanging holes also at the back of the unit. It is important to check 
that the 433MHz (if using wireless connection) and the DCF77 radio 
controlled time signal can be received before permanently mounting 
any of the units. Should the base station not display the 433MHz 
weather data from the sensors or the radio controlled time from the 
desired location, then relocate the units. Once the signals are re-
ceived, the system can be affixed. Also if you have selected to use 
cable connection, ensure that distances can reach all desired loca-
tions before affixing any unit permanently 
 
Mounting the Wind Sensor onto a mast  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Firstly, check that the wind-fan and the wind-vane can rotate freely 
before fixing the unit.  For correct and accurate readings it is impor-
tant to mount the sensor so that the front (marked E) is pointing in 
East-West direction.  The wind sensor should now be mounted using 
the screw provided onto a mast to allow the wind to travel around the 
sensor unhindered from all directions (ideal mast size should be from 
Ø16mm – Ø33mm).  

 

Once the wind sensor is fixed onto the mast, connect the cable to the 
corresponding thermo-hygro sensor socket so that operating power 
supply can be received and data can be transmitted to the base sta-
tion. 

 
 

Mounting the Rain Sensor 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
For best results, the rain sensor should be securely mounted onto a 
horizontal surface about 1 meter above the ground and in an open 
area away from trees or other coverings where rainfall may be re-
duced causing inaccurate readings. 

 

When securing into place, check that rain excess will not collect and 
store at the base of the unit but can flow out between the base and 
the mounting surface (test by pouring clean water). 

 

After mounting the rain sensor, connect the cable to the thermo-hygro 
sensor at the corresponding socket so power supply can be received 
and data be transmitted to the base station 

 

The rain sensor is now operable. For testing purposes, very slowly 
pour a small amount of clean water into the rain sensor funnel. The 
water will act as rainfall and will be received and displayed at the 
base station after about 2 minutes delay i.e. when the reading interval 
is reached (to clear this testing data on the base station, refer to the 
section “MIN/MAX Mode” below). 
 
 
 
 
 
 

Base portion 

Funnel portion 

 

Wind-fan 

Wind-vane 

Mast 

 

Summary of Contents for WS2307

Page 1: ...Features of the base station Features of the thermo hygro sensor Features of the wind sensor Features of the rain sensor 3 Safety Notes 4 Packaged contents 5 Setting up 6 Operation using cable connec...

Page 2: ...ther station is designed to work easily with your PC There are tedious requirements to turn your PC ON or OFF simply connect and disconnect the PC cable at any time System Requirements for PC use The...

Page 3: ...liable For reasons of safety and operation alterations to this device are strictly prohibited To operate the weather station use only supplied adaptor and batter ies of the recommended type Do not le...

Page 4: ...of the thermo hygro sensor to reveal the three sockets for the wind sensor rain sensor and the base sta tion 2 Connect the attached cables of wind and rain sensors to the corresponding sockets of the...

Page 5: ...and will result in higher power consumption Therefore batteries will have a shorter life span for cable connection compared to using 433MHz To operate using cable connection simply use the enclosed 1...

Page 6: ...door temperature and humidity 2 Outdoor temperature and humidity 3 Outdoor wind chill 4 Dew point 5 Rainfall 24 hour 6 Rainfall 1hour 7 Rainfall Total 8 Relative and absolute pressure 9 Wind speed win...

Page 7: ...ndered from all directions ideal mast size should be from 16mm 33mm Once the wind sensor is fixed onto the mast connect the cable to the corresponding thermo hygro sensor socket so that operating powe...

Page 8: ...C use please see PC user manual in the enclosed Heavy Weather CD ROM However if you wish to make a full reset of the base station and return to the original factory settings simultaneously press and h...

Page 9: ...ion will register this as a change in weather Air Pressure LCD Section 1 The air pressure reading is displayed here Press the PRESSURE key to toggle between relative and absolute air pressure displays...

Page 10: ...ection will be displayed on the LCD compass on the wind section Press the WIND key to toggle between wind direc tion as numerical e g 225 and abbreviated characters e g SW as well as numerical wind sp...

Page 11: ...nual setting mode allows the user to change several basic settings which is done by accessing one mode after the other simply by pressing the SET key After the final mode or if no key is pressed for 3...

Page 12: ...or min with time and date Rainfall 24 hours max or min with time and date Rainfall 1 hour max or min with time and date Rainfall total max only with time and date When in any of the above modes press...

Page 13: ...oggle to the indoor temperature low alarm set mode 5 Press and hold the SET key to enter the indoor temperature low setting values temperature digits will start flashing and set the desired indoor tem...

Page 14: ...will start flashing and set the desired wind chill low by using the PLUS or MINUS keys 6 Press ALARM key to confirm and press the MIN MAX key to re turn the normal display mode or press the ALARM onc...

Page 15: ...rm 2 Continue to press the ALARM key until you reach the wind direc tion alarm set mode 3 Press and hold the SET key to enter the wind direction setting values 4 Using the PLUS or MINUS keys select th...

Page 16: ...Tempera ture 30 0 C 69 9 C Indoor Temperature 9 9 C No alarm will sound if the minimum indoor temperature alarm is set below 9 9 C 59 9 C Humidity all 20 RH 95 RH Rainfall 24h 0 0 mm 1000 mm Rainfall...

Page 17: ...y in the thermo hygro sensor 1 Open the battery cover 2 Remove the old batteries and insert with new ones of the rec ommended type and replace the cover Once the sensor is powered up press and hold th...

Page 18: ...ese devices allow the change to an interference free frequency Such measures will effectively overcome interferences 20 Transmission Range The transmission distance from the thermo hygro sensor to the...

Page 19: ...seconds Indoor data Pressure temperature 4 times per minute Indoor Temperature Range 9 9 C to 59 9 C shows OFL if outside range Resolution 0 1 C Measuring Range Rel Humidity 20 to 95 if the relative...

Page 20: ...n of Conformity We hereby declare that this wireless transmission device does comply with the essential requirements of R TTE Directive 1999 5 EC EJIN23000012 For use in Germany Austria UK France Belg...

Reviews: