background image

11 

Sr. 

No. 

Function 

Send command 

Return command 

Remarks 

13 

Write device 

address 

01 06 00 19  00 02 D9 CC  01 06 00 19 00 02 

D9 CC 

Set to 2 

14 

Write baud 

rate 

01 06 00 1A 00 00 A8 0D  01 06 00 1A 00 00 

A8 0D 

Set to 2400 

15 

Write restore 

default 

01 06 00 1B 00 FF B9 8D  01 06 00 1B 00 FF 

B9 8D 

Factory default values 

are restored after sent 

16 

Write manual 

temperature 

01 10 00 3A 00 02 04 00 

00 41 A0 40 EC 

01 10 00 3A 00 02 

61 C5 

Set to 20.0 

17 

Write zero 

calibration 

01 06 00 3E 00 FF A8 46  01 06 00 3E 00 FF 

A8 46 

Confirm to calibration 

zero 

18 

Write slope 

calibration 

01 06 00 3F 00 FF F9 86  01 06 00 3F 00 FF 

F9 86 

Confirm to calibration 

slope 

19 

Read  

measuring AD 

01 03 00 66 00 01 64 15  01 03 02 2E E0 A4 

6C 

2EE0 turns to integer 

12000 

Maintenance And Storage 

 

The electrode should be cleaned regularly.  

 

Be careful while disassembling and rinsing the chlorine membrane,  

 

Do not wipe chlorine membrane on the electrode with filter paper or sandpaper. 

 

If the membrane head is fouled and clogged, the electrolyte is dry, lack or is  
contaminated. Stop using the membrane and its head should be removed and cleaned. 

 

After cleaning the electrode, replace the membrane head, add electrolyte. After  
long-term storage, it needs to be polarized and calibrated before use. 

 

The cable connector must be kept clean, dry, free from moisture, water, acids, alkalis or 
salts, etc. 

 

If water supply is cut off or electrode is not used on site, the electrode should be taken 
out, cleaned and covered with protective sleeve containing water for preservation.  

 

If the electrode is not used and stored for a long time, remove the electrode, disconnect 
the cable, drain the electrolyte, thoroughly clean the anode and the cathode with  

deionized water at 30 °C-40 °C, dry it and put on the protective cover. Place it in a dry 

place for storage. 

Summary of Contents for SRWQ100-YL105-9001

Page 1: ...ongevity please take a moment to familiarize yourself with this manual before powering the device Please keep it handy for future reference In case of any questions regarding the installation or use o...

Page 2: ...Make sure no bubbles are present in membrane head for accurate results Extension cutting joint and forcefully pulling signal line may cause inaccurate results Electrode Wiring Please follow the instru...

Page 3: ...is replaced every 18 to 24 months Follow the below procedure to replace the membrane head and add electrolyte Disconnect the power supply Turn the membrane head counter clockwise Remove the membrane...

Page 4: ...re shipment user can directly use it Online monitoring of the measured medium should keep a constant flow rate range 15 30 L h The calibration of dissolved oxygen electrode is performed by zero oxygen...

Page 5: ...ion Code Description This product supports 03 06 16 and other common function codes The output register uses 16 function codes when writing double word data or writing multiple data in batches Baud Ra...

Page 6: ...37 0x41 0xDB 0x8E 0x20 Slave Computer Response Format Note 72 37 41 DB transfer to floating point number CDAB change order is ABCD i e 41 DB 72 37 transfer to floating point is 27 4 Read Integer Host...

Page 7: ...omputer Transmission Format Slave Computer Response Format Note Change the local computer address 1 to address 2 and write the hexadecimal number 0X00 02 into register 0X00 19 ID Address Function Code...

Page 8: ...nt data to the 0x24 register Floating point 1 0 to hexadecimal number is 3F800000 high and low transposition is 00003F80 Send command 0110 00 24 00 02 04 00 00 3F 80 E0 14 The First Point Zero Calibra...

Page 9: ...00 12 Floating Point 2 Adjust measurement Temperature Offset 0x 00 14 Floating Point 2 Adjust temperature value Damping Coefficient 0x 00 16 Integer 1 0 10 Device Address 0x 00 19 Integer 1 1 255 Bau...

Page 10: ...6 Write upper limit of measurement 01 10 00 0A 00 02 04 00 00 41 20 42 58 01 10 00 0A 00 02 61 CA The upper measurement limit is set to 10 00 7 Write lower limit of measurement 01 10 00 0C 00 02 04 0...

Page 11: ...regularly Be careful while disassembling and rinsing the chlorine membrane Do not wipe chlorine membrane on the electrode with filter paper or sandpaper If the membrane head is fouled and clogged the...

Page 12: ...Modbus master device set tings and verify that the Modbus master device parity check is set to None Rs232 or RS485 cable is faulty Replace repair cables No network offsets and terminations or network...

Reviews: