11
Sr.
No.
Function
Send command
Return command
Remarks
13
Write device
address
01 06 00 19 00 02 D9 CC 01 06 00 19 00 02
D9 CC
Set to 2
14
Write baud
rate
01 06 00 1A 00 00 A8 0D 01 06 00 1A 00 00
A8 0D
Set to 2400
15
Write restore
default
01 06 00 1B 00 FF B9 8D 01 06 00 1B 00 FF
B9 8D
Factory default values
are restored after sent
16
Write manual
temperature
01 10 00 3A 00 02 04 00
00 41 A0 40 EC
01 10 00 3A 00 02
61 C5
Set to 20.0
17
Write zero
calibration
01 06 00 3E 00 FF A8 46 01 06 00 3E 00 FF
A8 46
Confirm to calibration
zero
18
Write slope
calibration
01 06 00 3F 00 FF F9 86 01 06 00 3F 00 FF
F9 86
Confirm to calibration
slope
19
Read
measuring AD
01 03 00 66 00 01 64 15 01 03 02 2E E0 A4
6C
2EE0 turns to integer
12000
Maintenance And Storage
The electrode should be cleaned regularly.
Be careful while disassembling and rinsing the chlorine membrane,
Do not wipe chlorine membrane on the electrode with filter paper or sandpaper.
If the membrane head is fouled and clogged, the electrolyte is dry, lack or is
contaminated. Stop using the membrane and its head should be removed and cleaned.
After cleaning the electrode, replace the membrane head, add electrolyte. After
long-term storage, it needs to be polarized and calibrated before use.
The cable connector must be kept clean, dry, free from moisture, water, acids, alkalis or
salts, etc.
If water supply is cut off or electrode is not used on site, the electrode should be taken
out, cleaned and covered with protective sleeve containing water for preservation.
If the electrode is not used and stored for a long time, remove the electrode, disconnect
the cable, drain the electrolyte, thoroughly clean the anode and the cathode with
deionized water at 30 °C-40 °C, dry it and put on the protective cover. Place it in a dry
place for storage.