HANDY DATA LOGGER: SME-100A/101A
1)
Number of measuring CHs
1 (When using the single SME series)
NTB enables sequential measurement in a maximum 33 CHs
*MAX. 33 CH
(SME series 1 CH + 8 units of the NTB (32 CHs)
2)
Sampling period
(When using the single SME series.
Varies with the connecting NTB.)
Approx. 0.5 sec:
0 to ± 30000 um/m
Approx. 1 sec:
± 30000 to ± 300000 um/m
Measurement with civil engineering
transducer with temperature measuring
function
3)
Measuring function
RELATIVE measurement (relative value measurement):
Each value is obtained by deducting the ZERO value.
*ZERO value is equivalent to the initial unbalance value.
Capable of obtaining the ZERO value at arbitrary timing.
4)
Coefficient calculation function
None
5)
Applicable sensor
Strain gage
Strain gage transducer
Civil engineering transducer with temperature measuring
function
Gage method
Applicable gage resistance
Quarter bridge
120, 240, 350 ohm
Half or full bridge
120 to 1000 ohm
6) Bridge
excitation
Constant-voltage bridge excitation:
Approx. 2 VDC
Constant-current bridge excitation:
(Bridge resistance: 350 ohm)
Approx. 5.6 mA
7) Measuring
range
Strain:
0 to ±300000 um/m (Constant-voltage bridge excitation)
0 to ±20000 um/m (Constant-current bridge excitation)
Temperature measurement with civil engineering transducer
with temperature measuring function: -30.0 to 70.0 °C
8) Resolution
Strain:
0 to ±30000 um/m: 1 um/m
±30000 to ±300000 um/m: 10 um/m
Temperature measurement with civil engineering transducer
with temperature measuring function: 0.1 °C
9) Accuracy
(when connected with one-touch
NDIS connector)
Strain:
0 to ±30000 um/m: ±(0.05% rdg + 2) um/m
±30000 to ±300000 um/m: ±(0.1% rdg +20) um/m
Temperature: ±0.5 °C
10) Check function
Insulation resistance measurement: 2 M to 100 Mohm
Resistance measurement: 0 to 20 kohm