background image

Summary of Contents for IAQ MAX

Page 1: ......

Page 2: ...3 Considerations 3 Product Specifications Content 4 Start Up Instructions 5 RPH 6FUHHQ LVSOD 6 CO2 Table Display 7 Settings Display 8 Charging 9 Calibration 9 Data Log Download Procedure 10 12 Mainten...

Page 3: ...real time NDIR CO2 sensor for fast accurate measurements Visual alarm indication Built in data log display table and downloadable software Fresh air calibration Powered by USB or Rechargeable Lithium...

Page 4: ...85 Barometric Pressure BARO 860hpa 1060hpa Storage Temperature 14 F to 140 F Data Logging Record 10 min intervals default Power source Lithium battery with 3000 mAh capacity Battery Life 5 hours 5V DC...

Page 5: ...to confirm highlighted options 5 When you hold down the center power button the air quality monitor will boot up The IAQ MAX detector will proceed through its warm up sequence for about 3 minutes to a...

Page 6: ...e current temperature level 3 Humidity HUM display area showing the current humidity level 4 Barometric Pressure BARO display area showing the current air pressure level CO2 Indoor Air Quality Grade R...

Page 7: ...ric Pressure BARO are shown in addition to a table that displays the last hour of CO2 readings The table updates on a 10 minute interval In order to download a comprehensive data set for further analy...

Page 8: ...e the and arrow buttons to adjust the highlighted setting It will save each setting automatically SET UP MENU DATE User set date TIME User set time UNIT Choose F or C for Temperature INVL Data logging...

Page 9: ...ing with a USB computer port which only outputs 500mA as this will provide a much slower charge 11 Calibration The IAQ MAX has two different CO2 calibration methods 1 Auto Calibration calibrated using...

Page 10: ...AQ Series and MAX Model from the GasLab Software drop downs under Sensor Select and Click CONNECT 12 Data Logging Setup The device will begin to data log upon power up The data logging interval can be...

Page 11: ...13 Data Log Download Procedure Continued 4 Click on Configure Sensor 5 Click on Download Datalog Name and Save the File appropriately and Press OK when prompted 11...

Page 12: ...6 Finally View real time data analysis 13 Data Log Download Procedure Continued 7 Locate and open your saved file for further analysis This is an example below of the exported data set 12...

Page 13: ...please observe the following guidelines 1 Repair Do not attempt to repair or modify the device in any way Please contact a CO2Meter specialist directly if the product needs servicing including replac...

Reviews: