3-6 SENSOR CONNECTION
For the strain gage transducer with a connector plug (NDIS standard) applied to the cable tip, directly
connect the transducer to the SME series connector. For the strain gage or transducer without the connector
plug, use the accessory input cable for connection. Wirings with the input connector or input cable are
described as follow.
1) Input
connector
Measuring Target
Connection Method
Remarks
Strain gage
A:
+
BV
Strain gage transducer
B:
-
SIG
Civil engineering transducer with temperature
C:
-
BV
14
measuring function
Quarter bridge method
120,240,350 ohm
Full bridge method
120 to 1000 ohm
D:
+
SIG
E: Shield
F: TEDS
+
G: TEDS
-
TEDS compatible
2) Input
cable
Bridge Circuit
: Strain gage
: Built-in fixed
Measuring Method
resistor
A: Active gage
E: Bridge exciter
e: Bridge output
Connection Method
A: Active gage
CH mode
No measurement.
SKIP
Measuring data is space.
Quarter bridge method
(2-wire system connection)
Gage resistance
120, 240, 350 ohm
1G120
1G240
1G350
Red (+BV)
White (-SIG)
Black (-BV)
Green (+SIG)
A
Crocodile clip
(Shield)
Quarter bridge method
(3-wire system connection)
Gage resistance
120, 240, 350 ohm
1G120
1G240
1G350
Red (+BV)
White (-SIG)
Black (-BV)
Green (+SIG)
A
Crocodile clip
(Shield)