40
DPM-911/951
DPM-912/952
DPM-913
S/N Ratio
when 500 m/m is input,
outputs 10.00 V.
911 54 dBp-p or more
(RTI: Within 2 m/mp-p)
951 53 dBp-p or more
(RTI: Within 2.3 m/mp-p)
912 52 dBp-p or more
(RTI: Within 2.5 m/mp-p)
952 47 dBp-p or more
(RTI: Within 4.5 m/mp-p)
48 dBp-p or more
(RTI: Within 3.9 m/mp-p)
when 1000 m/m is input,
outputs 10.00 V.
911 60 dBp-p or more
951 58 dBp-p or more
912 57 dBp-p or more
952 53 dBp-p or more
53 dBp-p or more
[Common Condition] Excitation Voltage: 2 Vrms,
Bridge Resistance: 120 ohm, LPF = FLAT
Output
OUTPUT A: ±10 V (Load resistance 5 kohm or more)
OUTPUT B: ±10 V (Load resistance 5 kohm or more)
Stability
Zero - Temperature
±0.1 m/m/ºC
±0.2 m/m/ºC
Zero - Time
±0.5 m/m/24h
±1.0 m/m/24h
Zero – Power Fluctuation
±0.05% FS/±10% fluctuation
±0.05% FS/±10% fluctuation
Sensitivity - Temperature
±0.05%/ºC
±0.05%/ºC
Sensitivity - Time
±0.3%/24 h
±0.3%/24 h
Sensitivity – Power Fluctuation
±0.05%/±10% fluctuation
±0.05%/±10% fluctuation
Withstand Voltage
Between measuring bridge – case : 1000 VAC for 1 minutes
Between AC power – case : 1000 VAC for 1 minutes
Output Voltage Display
4
1
/
2
-digit digital display (7-segment LED)
11-segment LED bar graph meter
Excessive Input Display
Output Voltage Display flickers. (4
1
/
2
-digit digital display only)
Check Function
Bridge Check
Key-lock Function
Setting methods other than POWER switch become invalid
(However, the setting values of the CAL and VOLTAGE OUT switches
can be changed.)
Remote Function
Capable of controlling the following functions
Balance adjustment (BAL)
Addition of the calibration signal (CAL)
Key-lock
Synchronization Method
Automatically determines internal (INT) or external (EXT)
and manual setting
TEDS
Read the TEDS information of the TEDS-compatible sensor and set the
rated output to the output voltage of the VOLTAGE OUT switch
(Condition: Within the setting range of the sensitivity adjuster)
Actual Load Calibration
Set the actual load to the output voltage of the VOLTAGE OUT switch
(Condition: Within the setting range of the sensitivity adjuster)
Vibration Resistance
5 to 200 Hz, 29.4 m/s
2
(3G)
X, Y, and Z directions: 12 cycles with 10 minutes/cycle
Impact Resistance
15 G, 11 msec or less
X, Y, and Z directions: each 3 times
Operating Temperature &
Humidity Range
-10 to 50ºC, 20 to 85 %RH (noncondensing)
Storage Temperature Range
-30 to 70ºC
Power Supply
90 to 110 VAC (approximately 10 VA: 100 VAC)
10.5 to 15 VDC (approximately 0.6 A: 12 VDC)
The following products are also available upon request.
A115: 108 to 132VAC
A200: 180 to 220VAC
A230: 207 to 253VAC
Dimension
49 (W) mm× 128.5 (H)mm × 262.5 (D) mm (Protruded portion not included)
Weight
Approximately 1.2 kg