5-15 TPN/DLT MEASURING MODE
(
For the DPM-911/951 only
)
In addition the strain gages or strain gage transducers, the DPM-900/950 series can measure data using
Rotary Transformer Torque Transducer, and Inductance-Type Displacement Transducer.
Be sure to change modes to the TPN/DLT Measuring Mode for using the Rotary Transformer
Torque Transducer or Inductance-Type Displacement Transducer.
To change modes to the TPN/DLT Measuring Mode, turn ON the Setting Switch 4 on the rear panel. Be
sure to turn OFF the DPM-900/950 series before changing the Setting Switches.
In addition, for how to set the internal gain, read 5-7 INTERNAL GAIN.
Turn ON the “Setting Switch 4” and immediately after turning ON the power, check that the monitor
display changes as follows.
All LED light up ® Version of the built-in software ® ‘
’
(
Scroll horizontally
)
® Output voltage
The
appears in the TPN/DLT Measuring Mode only.
Do not conduct the bridge check function with the Rotary Transformer Torque Transducer
connected.
If the “Setting Switch 4” of the DPM-912/913/952 is turned ON, the DPM-912/913/952 is still in
the Normal Mode.
NOTE
After turning ON the DPM-900/950 series, press and hold the
SET
switch for checking the TPN/DLT
Measuring Mode.
The following display appears on the monitor.
The version of the built-in software and are scrolled horizontally.
Press and hold the
SET
switch if the Setting Switch 4 is set to OFF.
The following display appears.
The version of the built-in software is scrolled horizontally.