5-12 APPLICATION OF CENTRALIZED CONNECTOR
The DPM-900/950 series provides a centralized connector on its rear panel. Generally, the connector is
used to connect with an AC power supply via the accessory power cable. The connector also provides a pin
that connects with a DC power supply. The connector also enables connection of input/output signals.
Names and functions of these signals are shown in the following table.
For details of external control signals (with REM on the signal name) in the table, read “5-13 REMOTE
CONTROL.”
Pin No.
Signal Name
Signal type
Function and destination
1
+12 V
Power supply
Connects
1
with the + terminal of a DC power
supply.
2
REM BAL
2
Remote input
Valid with an external balancing signal “L.”
3
OSC
Input/output
Connects with input synchronous signal.
4
N.C
3
5
COM
Common electric
potential
Common electric potential of the remote signal,
bridge status, OUTPUT A
Also can be used as OCS – signal.
6
N.C
3
7
AC POWER INPUT
Power supply
Connects with an AC power supply.
8
N.C
3
9
Bridge status
Output
Outputs open collector
*12
“L” against a bridge
check which has proven N.G.
10
REM KEYLOCK
2
Remote input
Valid with an external key-lock signal “L.”
11
GND
Grounding terminal Connects with the case and COM terminals
internally.
12
N.C
13
N.C
14
0 V
Power supply
Connects with the minus terminal of a DC power
supply.
1
15
OUTPUT A
2
Output
Connects with OUTPUT A connector internally.
16
COM
Common electric
potential
Common electric potential of the remote signal,
bridge status, OUTPUT A
Also can be used as OCS – signal.
17
REM +CAL
2
Remote input
Valid with an “L” signal that commands execution
of exCAL input.
18
REM -CAL
2
Remote input
Valid with an “L” signal that commands execution
of external -CAL input.
19
N.C
3
20
100 VAC
Power supply
Connects with an AC power supply.
Connector plug: HIROSE ELECTRONIC S-1620A-C
1
Use a heat-resisting vinyl cable of over 0.3 mm
2
in diameter for wiring.
2
It is a signal with Pin #5 COM as common electric potential. (COM and GND have same electric
potential)
3
N.C: Non connection
5-13 REMOTE CONTROL
The following functions can be controlled remotely.
Execution of balancing
REM BAL signal
Outputting CAL
REM +CAL, REM –CAL signals
Execution and cancellation of key-lock
REM KEYLOCK signal
Remote control has the same significance as manual operation of the front panel switches that control the
DPM-900/950 series functions. The priority is given to the first operated one.