Frequency Hz ( AC Current )   (Auto-range)

Range

Display range

Allowable input

Accuracy

10Hz

0.000

9.999Hz

The accuracy is not guaranteed

100Hz

9.00

99.99Hz

9.00Hz

400.0Hz

±0.2

rdg±2dgt

1000Hz

90.0

400.0Hz

±0.1

rdg±1dgt

400.1

999.9Hz

The accuracy is not guaranteed

10kHz

0.900

9.999kHz

100kHz

9.00

99.99kHz

1000kHz

90.0

999.9kHz

10MHz

0.900

9.999MHz

Input Current

more than 6A

Frequency Hz ( AC Voltage )    (Auto-range)

Range

Display range

Allowable input

Accuracy

10Hz

0.000

9.999Hz

The accuracy is not guaranteed

100Hz

9.00

99.99Hz

9.00Hz

300.0kHz

±0.2

rdg±2dgt

1000Hz

90.0

999.9Hz

±0.1

rdg±1dgt

10kHz

0.900

9.999kHz

100kHz

9.00

99.99kHz

300kHz

90.0

300.0kHz

1000kHz 300.1

999.9kHz

The accuracy is not guaranteed

10MHz

0.900

9.999MHz

 Input  Voltage

more than 6V (

10kHz) , more than 20V (10k

300kHz)

Input Impedance

approx. 900k

Ω

Note:

 The symbol of "

" in the above table means that the 

instrument only displays the value, but the accuracy, the 

proper operation and the safety are not guaranteed. 

Safety Standard 

IEC 61010-1, 61010-2-032, 61010-2-033

 

CAT III 300V, pollution degree 2

 

CAT II 600V, pollution degree 2

 

IEC 61010-031

 

EN 61326-1 (EMC)

 

EN 50581 (RoHS)

Operating System 

ΔΣ

 modulation

Display 

Liquid crystal display

 

Maximum Reading:6039

 

Except Hz : 9999, CAP : 4039, 

 

Diode : 1999

 

Bar graph with maximum points of 30.

Display 

approx. 3 times per second

Indication renewal

Location for use 

 Indoor use, 2000m max, above sealevel

Operating Temperature   0 ~ +40

, relative humidity 85% or less

and Humidity Range 

 

(without condensation)

Storage Temperature  -20 ~ +60

, relative humidity 85% or less

and Humidity Range 

 

(without condensation)

Source 

Two 1.5VDC R03 (UM-4) batteries

Current Consumption   approx. 3mA (DCV), approx. 13mA 

(ACA)

Power-save Function   Shifts to the power-save state about 

15 minutes after the last switch 

operation.

Low battery warning 

   

  Appears when the batteries 

become low (2.4±0.15V or less)

Overload Protection 

 AC voltage / DC voltage / Frequency 

ranges : 

 

DC / ACrms 720V for 10 seconds

 

AC current / DC current ranges : 

 

DC / ACrms 150A for 10 seconds

 

 Resistance / Continuity / Diode / 

Capacitor ranges : 

 

DC / ACrms 600V for 10 seconds

Withstand Voltage 

 AC3470Vrms for 5 seconds between 

electrical circuit and housing case

Insulation Resistance  100M

Ω

 or greater at 1000V 

 

 between electrical circuit and housing 

case

Conductor Size 

Approx. 

φ

12mm diameter max

Dimensions 128(L)×92(W)×27(D)mm

Weight 

Approx. 220g

Accessories 

Two R03 (UM-4) batteries

 

Instruction Manual

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for KEW MATE 2012RA

Page 1: ...DIGITAL MULTIMETER WITH AC DC CLAMP SENSOR KEW MATE 2012RA www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...his instrument has been designed and tested according to IEC Publication 61010 Safety Requirements for Electronic Measuring Apparatus This instruction manual contains warnings and safety rules which m...

Page 3: ...omes standard with the instrument Clamp sensor for ease of use in crowded cable areas and other tight places Permits current measurement with an open current clamp sensor that does not require opening...

Page 4: ...Input Voltage 600V Range Display range Allowable input Accuracy 600mV 0 0 603 9mV 0 0m 600 0V 1 0 rdg 3dgt 6V 0 560 6 039V 60V 5 60 60 39V 600V 56 0 603 9V Input Impedance approx 10M Resistance Auto...

Page 5: ...g System modulation Display Liquid crystal display Maximum Reading 6039 Except Hz 9999 CAP 4039 Diode 1999 Bar graph with maximum points of 30 Display approx 3 times per second Indication renewal Loca...

Page 6: ...the batteries according to section 8 Battery Replacement NOTE When the instrument is left powered on the auto power save function automatically shut the power off The display blanks even if the Functi...

Page 7: ...LCD 2 Adjust one of the conductors to the center of the clamp sensor s arrow When the position of the conductor is not at the center of the arrow the error occurs Measured value is shown on the displa...

Page 8: ...urement DANGER In order to avoid possible shock hazard never make measurement on circuits with a maximum voltage difference of 600V or greater between conductors 300V or greater between a conductor an...

Page 9: ...Turn the function switch to the OFF and then to any position 7 2 Data Hold Function This is a function to freeze a measured value on the display Press the Button once to hold the current reading In t...

Reviews: