● Standards
IEC61010-1 Measurement CAT.
Ⅲ
300V, pollution degree 2 / Measurement CAT.
Ⅱ
600V, pollution degree 2
IEC61010-031
IEC61326
● Mode of operation
△∑ mode
● Indication
LCD maximum value 6000 (ACV/A, DCV/A, Ω) / maximum value 9999 (Hz) / maximum valve 4000 (F) , units, symbols
● Over range display
"OL" symbol is displayed on the LCD.
"OL" displayed when exceeding the measuring ranges. (except for AC/DC 600V ranges and AC/DC 10A ranges.)
● Auto-ranging
Range shifts to upper range when indicated value is more than 6040.
Range shifts to lower range when indicated value is less than 560.
● Sampling rate
approx. 400ms
● Operating Environmental
−indoor use
conditions
−altitude up to 2000m
● Temperature & Humidity range
(guaranteed accuracy)
23℃±5℃
Relative humidity: less than 75%
● Operating Temperature &
Humidity range
0℃〜+40℃
Relative humidity: less than 80%
● Storage Temperature &
Humidity range
-20℃〜+60℃ Relative humidity: less than 90%
● Insulation Resistance
It should be more than 100MΩ/DC1000V between electrical circuit and enclosures.
● Withstand Voltage
It should be more than AC3700V/ for one minute between electrical circuit and enclosures.
● Overload Protection
Voltage function
:
720V (RMS.) 10seconds
Resistance function
:
600V (RMS.) 10seconds
Capacitance function
:
600V (RMS.) 10seconds
Frequency function
:
600V (RMS.) 10seconds
Current function : uA, mA
:
Protected by 600V 0.8A fuse
A
:
Protected by 600V 10A fuse
● Dimensions / Weight
approx. 161(L) x 82(W) x 50(D) mm / approx. 280g(including batteries, holster installation)
● Power source
Two R6P(AA)1.5V or equivalent
● Accessories
Test lead 1set / R6P (AA) 2p.c.s / Holster 1 / Instruction manual 1
● Fuse
F600V/800mA (fast acting type), φ6.3 x 32mm / F600V/10A(fast acting type), φ6.3 x 32mm
CAUTION
● The voltage shown above is the overload protection(overvoltage protection)for the instrument. Make sure not to exceed the value of voltage shown above.
● In locations subject to radio frequency interference, the product may malfunction and it will recover automatically after the test.
― 15 ―