To reduce the risk of fire or electric shock, do not expose this product
to rain or moisture. The meter is intended only for indoor use.
To avoid electrical shock hazard, observe the proper safety
precautions when working with voltages above 60 VDC or
30VAC rms. These voltage levels pose a potential shock hazard to
the user. Before & after hazardous voltage measurements, test the
voltage function on a known source such as line voltage to determine
proper meter functioning.
Identifies conditions and actions that could result
in serious injury or even death to the user.
Identifies conditions and actions that could cause
damage or malfunction in the instrument.
Keep your hands/fingers behind the hand/finger barriers (of the
meter and the test leads) that indicate the limits of safe access of
the hand-held part during measurement. Inspect test leads,
connectors, and probes for damaged insulation or exposed metal
before using the instrument. If any defects are found, replace them
immediately.
Only use the test lead provided with the equipment or
UL Listed Probe Assembly rated CAT IV 1000V or better.
Disconnect the test leads from the test points
before changing meter functions.
This Clamp-on meter is designed to apply around or remove from
uninsulated hazardous live conductors. But still, individual protective
equipment must be used if hazardous live parts in the installation
where measurement is to be carried out could be accessible.
- 02 -
1)
This manual contains information and warnings that must be
followed for operating the instrument safely and maintaining the
instrument in a safe operating condition. If the instrument is
used in a manner not specified by the manufacturer, the
protection provided by the instrument may be impaired.
Measurement Category IV 1000V AC & DC.
OVERVOLTAGE CATEGORY ll (CAT II)
is for equipment
intended to be supplied from the building wiring. It applies both to
p l u g - c o n n e c t e d e q u i p m e n t a n d t o P E R M A N E N T LY
CONNECTED EQUIPMENT.
The meter protection rating, against the users, is double
insulation per UL/IEC/EN61010-1 Ed. 3.0, IEC/EN61010-2-033
Ed. 1.0, CAN/CSA C22.2 No. 61010-1 Ed. 3.0, IEC/EN61010-2-
032, Ed. 3.0 & IEC/EN61010-031 Ed. 1.1:
OVERVOLTAGE CATEGORY lV (CAT IV)
is for equipment
installed at or near the origin of the electrical supply to a building,
between the building entrance and the main distribution board.
Such equipment may include electricity tariff meters and primary
overcurrent protection devices.
Per IEC61010-1 OVERVOLTAGE CATEGORY
OVERVOLTAGE CATEGORY lll (CAT III)
is for equipment
intended to form part of a building wiring installation. Such
equipment includes socket outlets, fuse panels, and some
MAINS installation control equipment.
- 01 -