3)  SPECIFICATIONS :

GENERAL SPECIFICATIONS :

SPECIAL FEATURES:

 

  

Extended >1000 ~ 1500V DC range suitable for solar

  

TM

AutoCheck   Voltage & Ohms

  

VFD-V & VFD-Hz function

  

5ms CREST-MAX capture mode (Peak Hold)

Autoranging Relative -Zero mode

  

  

Display Hold function & 

EF-Detection (NCV)

Backlight LCD Display  

  

Auto Power Off

  

Diode Test & Continuity Test

  

PC Interface (Optional) 

  

 

Sensing :

 TRMS sensing 

Jaw Opening  :

 55mm max.

 

Display :

 3-5/6 digits 6000 counts & 3½ digits 1,999 counts for Hz.

 

Update Rate :

 5 per second nominal

 

Polarity :

 Automatic

 

Low Battery :

 Below approx 2.4V

 

o

o

Operating Temperature :

 0 C to 40 C

 

 

Relative Humidity :

 Maximum  80%  R. H. for temperature upto 

o

o

    31 C decreasing linearly to 50% Relative Humidity at 40 C

o

o

Storage Temperature :

  -20 C to 60 C, < 80% R.H.   

 

     (With battery removed)

Altitude :

 Operating below 2000m

 

Temperature Coefficient :

 nominal 0.15 x (specified accuracy) / 

 

o

o

o

o

     °C @ (0 C--18 C or 28 C--40 C), or otherwise specified 

- 04 -

The instruments conform to CENELEC Low-Voltage directive 

2006/95/EC  and  Electromagnetic  compatibility  directive 

2004/108/EC.

Caution !  Refer to the explanation in this Manual

Caution !  Risk of electric shock

Earth (Ground)

Double Insulation or Reinforced insulation

Fuse

AC--Alternating Current

DC--Direct Current

Application around and removal from hazardous live 
conductors is permitted

2) CENELEC Directives

INTERNATIONAL ELECTRICAL SYMBOLS

- 03 -

Summary of Contents for KM 2778

Page 1: ...Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function Generator Phasing Sticks Battery Tester Waterproof Pen Tester...

Page 2: ...ce Extended 1500V DC Range SAFETY 01 TERMS IN THIS MANUAL 02 INTERNATIONAL ELECTRICAL SYMBOL 03 FEATURES GENERAL SPECIFICATION 04 ELECTRICAL SPECIFICATION 06 PRODUCT DESCRIPTION 14 OPERATION 15 MAINTE...

Page 3: ...uninsulated hazardous live conductors But still individual protective equipment must be used if hazardous live parts in the installation where measurement is to be carried out could be accessible 02...

Page 4: ...C Relative Humidity Maximum 80 R H for temperature upto o o 31 C decreasing linearly to 50 Relative Humidity at 40 C o o Storage Temperature 20 C to 60 C 80 R H With battery removed Altitude Operatin...

Page 5: ...onsumption Typical 14mA for Current function 5 2mA for others APO Timing Idle for 34 minutes APO Consumption 10 A typical Dimension 264 L x 97 W x 43 H mm Weight Approx 608gm Accessories Test leads pa...

Page 6: ...2 Accuracy Resolution Range True RMS Crest Factor 1 4 1 at full scale 2 8 1 at half scale 1 Induced error from adjacent current carrying conductor 0 1A A 2 Specified with Relative Zero mode applied to...

Page 7: ...0 00 F 2 3 5 rdg 5dgts Resolution Range 2 4 0 rdg 5dgts 1 Accuracy 600 0 F 2000 F 1 Accuracies with film capacitor or better 2 Temperature Coefficient 0 25 x specified accuracy C 0 C 18 C or 28 C 40 C...

Page 8: ...DC typical Open Circuit Voltage Accuracy 1 0 3d 11 TM AUTOCHECK _DCV Accuracy Resolution Range TM AutoCheck Lo Z DCV Threshold 1 5VDC 1 5VDC nominal TM AutoCheck Lo Z DCV input impedance Initially app...

Page 9: ...pick up 9 Jaw center DCA polarity Indicator at where best DCA ACA accuracy is specified 8 Jaw trigger for opening the clamp jaw 13 6 000 V 1 5 rdg 5dgts 60 00 V 600 0 V 1000 V 1 Accuracy Resolution Ra...

Page 10: ...users of being in such low impedance mode Peak initial load current while probing 1000VAC for example can be up to 566mA 1000V x 1 414 2 5k decreasing abruptly to approx 3 37mA 1000V x 1 414 420k TM w...

Page 11: ...ble Frequency Device signals It however further pre selects the most appropriate voltage ranges and thus the Hz trigger levels to best cope with most VFD Voltage and VFD Frequency applications Note Th...

Page 12: ...apacitance function in a live circuit will produce false results and may damage the meter In many cases the suspected component s must be disconnected from the circuit to obtain an accurate measuremen...

Page 13: ...through the test leads terminals Be sure to insert the banana plug type K temperature bead probe Bkp60 with correct polarities You can also use a plug adapter bkb32 optional purchase with banana pins...

Page 14: ...ature freezes the display for later view Press the HOLD button momentarily and release to toggle the hold feature PC computer interface capabilities Hold Press CREST HOLD button for one second or more...

Page 15: ...e resistance input terminal has subjected to high voltage transient caused by lightning or switching surge to the system under test by accident or abnormal conditions of operation the protective imped...

Page 16: ...78 The meter uses standard 1 5V AA Size IEC LR6 battery X 2 Loosen the 2 captive screws from the battery cover case Lift the battery cover case Replace the batteries Replace battery cover case Re fast...

Page 17: ...ected contaminated or damaged by accident or abnormal conditions of operation or handling All transaction are subject to Mumbai Jurisdiction KUSAM MECO authorized dealer shall extend this warranty on...

Reviews: