2. AC/DC measurement switch

3. DC Calibration

Before measuring DC current, press DCA ZERO key to
reduce  the  residual  magnetism  to  Zero,  and  then 
conduct  measurement.  Rational  usage  of  this  Adjust
Zero function will make the results more accurate. For
example, after boot, before measurement, we can take
the jaw close to the DC current wire. LCD will show an
inductive current. Press DCA ZERO to calibrate, which
deduct the inductive value, Showing below:

+

+

1+

1-

4. Current Measurement

Dangerous!  Please  don't  use  it  to  measure  current 
higher  than  600A;  otherwise  there  would  be  risk  of
electric shock to cause personal injuries or equipment
damages

!

High  voltage,  extremely  dangerous!  Operators  shall 
strictly keep to safety rules; otherwise there would be 
risk  of  electric  shock  to  cause  personal  injuries  or 
casualties.

The tester default to DC measurement after boot up. 
Press  ~/

Ì

  to  switch  the  AC/DC  function.  LCD  will 

display DC or AC to corresponding switch function.

4) Read LCD display data. In case 

OL mA 

symbol was

    displayed, it means that current of measured line is
    beyond the maximum limit of this meter.

1) Power on.
2) Press ~/

Ì

 to DC or AC measurement. 

    Conduct  Adjust Zero before measure DC current.

3) Release the clamp lever to open input clamp and clamp 
    measured conductors. Be attention that the clamp must 
    be  totally  closed  and  the  measuring  wire  passing 
    through the center of the jaw.

+

Wire Position

L

N

Measure AC line current

Measure consumption

Summary of Contents for KM 2009

Page 1: ...used when display FULL Carrying Case Battery installed User s Manual RS232 PC Interface CD ACCESSORIES Power Consumption about 10mW Peak Hold Press HOLD without release the meter will show the peak va...

Page 2: ...alysers Frequency Counter Function Generator Phasing Sticks Battery Tester Waterproof Pen Testers Solar Power Meter EMF Detector Wood Paper Grain Moisture Meter Transistorised Electronic Analog Digita...

Page 3: ...PMETER Model KM 2009 INDEX Warning 01 Introduction 02 Electrical Signs 03 Electrical Specifications 03 Clampmeter Structure 05 Operating Method 06 How to change Battery 11 Accessories 11 Test Certific...

Page 4: ...t safety operation based on instruction listed in this manual e g and danger signs on this manual Please use mA to test leakage current test by clamping grounding line single phase line together I INT...

Page 5: ...ay domain 32mm x 15mm 2 times s 50 60Hz Automatically DC current auto indentified and display Clamp Size Test Position Range Shift Line Voltage RS232 Interface Com Configure Data Memory 30mm x 35mm Te...

Page 6: ...ive polarity indication 4 LCD battery symbol 5 Data hold indication 6 Stored data code VI OPERATING METHOD 1 Start up Shutdown Press POWER key to start up LCD will begin to display Press the POWER key...

Page 7: ...or equipment damages High voltage extremely dangerous Operators shall strictly keep to safety rules otherwise there would be risk of electric shock to cause personal injuries or casualties The tester...

Page 8: ...to 60 groups press HOLD key again the FULL symbol will display which means storage memory is full press HOLD key to cancel FULL flickering and return to measuring mode I 2 Press HOLD POWER keys to en...

Page 9: ...out test when the battery cover plate is not in its position Please pay attention to the polarity of battery to avoid damaging the meter Change the low battery in time If not use the meter for a long...

Page 10: ...r shall extend this warranty on new and unused products to end user customers only but have no authority to extend a greater or different warranty on behalf of KUSAM MECO KUSAM MECO s warranty obligat...

Reviews: