Kurz Instruments Inc. 

 

8/18/2010 

 

 
360209-B Rev. B 

B-series Operation 

B - 3 

Program Mode accessed using the (P) key 
Display Mode accessed using the (D) key 
Log Mode accessed using the (L) key 
Extended Utilities Mode accessed using the (E) key 

 

Run Mode

 is the normal operational state of the meter.  After boot up or power on, the 

flow meter will be in 

Run Mode 

and will show the 

Run Mode

 display(s).  The factory 

default 

Run Mode

 display will show the meter's Tag Name and Flow as follows: 

 

TAG NATURAL GAS1 
FLOW 14356.  SCFM 

 
The 

Run Mode

 display can be configured to display other meter variables and formats.  

The 

Run Mode

 display setup

 is covered in a later section of the manual. 

 

Program Mode

 is the mode used to change (or program) the flow meter parameters or 

setup the flow meter features.  An access code is required to enter this mode.  

Program 

Mode

 is invoked by pressing 

P

 when the flow meter is in 

Run Mode

 

Display Mode

 is the mode used to view the flow meter parameters as well as the process 

and control data.  No access code is required to enter this mode.  This mode is for 
viewing only and data entry is blocked.  

Display Mode

 is invoked by pressing 

D

 when the 

flow meter is in 

Run Mode

 

Log Mode

 is the mode used to request log reports that are sent to the flow meter USB 

port.  No access code is required to enter this mode.  

Log Mode

 is invoked by pressing 

L

 

when the flow meter is in 

Run Mode

 

Extended Utilities Mode

 is the mode used to perform various diagnostic and utility 

functions.  An access code is required to enter this mode.  

Extended Utilities Mode

 is 

invoked by pressing 

E

 when the flow meter is in 

Run Mode

 
The 

H

 key provides a quick exit or ‘HOME’ function and will either return to the top-level 

option entry display of the user mode or return to 

Run Mode

 
The general format of the data displays will have the item description or name of the 
variable on line 1 of the LCD display and the value of the variable on line 2 of the 
display.  In general, data entry will be performed on the value on line 2 of the display.  
When the data item is a selection list, an up/down arrow ‘^v’ symbol will display at the 
right most position on line 2 to indicate that the data value is selected using the up 

(^)

 or 

down 

(v)

 arrow keys on the keypad.  In general, when the data item has associated 

engineering units, the units string will be at the right most position on line 2.   

Summary of Contents for MFT-B Series

Page 1: ...the Flow The Run Mode display can be configured to display other meter variables see section below If the cover is removed as shown in the picture the optional keypad is accessible to view change the...

Page 2: ...seconds for the whole boot process under flow For a full description of this and flow rate dependent boot times see this section The parameters to configure or setup the mass flow transmitter are acc...

Page 3: ...and data entry is blocked Display Mode is invoked by pressing D when the flow meter is in Run Mode Log Mode is the mode used to request log reports that are sent to the flow meter USB port No access c...

Page 4: ...nding on the user mode invoked the user will be prompted to enter an access code ENTER ACCESS CODE An access code is entered using the numeric keys on the keypad and pressing E to accept the entry The...

Page 5: ...ar with the Option s associated with the menu options the Quick Jump navigation allows the user the enter the option using the numeric keys and press E to quickly invoke the menu option The following...

Page 6: ...PID SETUP 22 Manual PID Adjust 22 PID CONTRL 23 Setup Drift Check 23 DRIFT CHCK 24 Change User Password 24 CHNG PWORD 25 Update FROM EEPROM 25 GET EEPROM 26 Set Bootup Output Delay 26 BOOTUP DLY The...

Page 7: ...rection Data 15 VRMS DATA 16 Remote Correction Factor Data 16 REMOTE CF 17 Select Gas Calibration Curve 17 CAL CURVE 18 Setup Data Logging 18 DATA LOG 19 Setup Modbus Communication 19 MODBUS COM 20 Se...

Page 8: ...and not changed LOG MODE MENU OPTIONS Option Menu 1 Event Log 1 EVENT 2 Min Max Data 2 MIN MAX 3 Trend Log 3 TREND 4 System Configuration 4 CONFIG 5 System Monitored Data 5 RUN DATA EXTENDED UTILITIE...

Page 9: ...st digit For parameters with a selection list the up arrow and down arrow v keys are used to scroll the list to the desired value When the desired list value is displayed the E key is pressed to accep...

Page 10: ...ho is off Note The meter Tag Name and other text parameters can be entered as a continuous string using a remote terminal keyboard versus single characters followed by the ENTER key by holding the SHI...

Page 11: ...8E 07 FLOW VEL FLOW 0 0000 SCFM VEL 0 00000 SFPM TAG FLOW TAG NAT GAS FLOW 0 0000 SCFM FLOW TEMP FLOW 0 0000 SCFM TEMP 82 003 DEGF If SCROLLED format is selected for the Run Mode Display the meter wil...

Page 12: ...value with units will be on line 2 Flow FLOW RATE 0 0000 SCFM Temp TEMPERATURE 82 52 DEGF Velocity VELOCITY 0 0000 SFPM Totalized Flow TOTAL FLOW 183545 0 SCF Tag Name TAG NAME NAT GAS24 For scrolled...

Page 13: ...changes in Program Mode the changes are automatically saved to EEPROM memory when the user exits from Program Mode Configuration Upload Download using Xmodem The upload download process allows the use...

Page 14: ...erminal echo ON and OFF Calibration of the Analog Output A menu is available to perform the calibration of the analog output s on MFT series flow meters You enter the Analog Output Calibration menu an...

Page 15: ...can be setup to energize a relay output when an alarm event occurs The solid state relays SSR are SPST and open circuit when the meter is unpowered The second alarm mechanism is the NE 43 Alarm Meter...

Reviews: