3 Cleaning and Maintenance
29
QuickGene-Auto12S/QuickGene-Auto24S Operation Manual
3
3.2 Weekly Maintenance
I.
Lubricate O-rings
i.
Apply grease entirely to O-rings by fingers.
ii.
Take off the excessive grease by paper
II.
(Optional) UV Irradiation. Please refer to page 18
i.
At “Home” screen, select “System”
ii.
At “System” screen, select “UV lamp”.
iii.
Make sure the front door already has been closed. Press “OK” button to start the UV
irradiation.
iv.
If it is necessary, you can press “Pause” button.
3.3 Half a Year Maintenance
It is recommended to replace O-ring every half a year.
(It may be changed up to your frequency of use.)
i.
Use a sharp blade to cut off old O-ring
ii.
Put grease on O-ring
iii.
Apply with new one
Please wipe the excessive grease by paper.