The electrode relay (K1) must meet the following requirements:
▪
Sensor circuit 10 to 30V ~
▪
Tripping current 0.5 to 3 mA
(equivalent to a tripping resistance of 3 to 60 k
Ω)
5.4.1.4.3 Leakage at the mechanical seal (optional)
K 1
7
6
bl
br
B2
B5
a)
b)
Fig. 6: a) Wiring of the electrode relay and b) Position of the leakage sensor
The submersible mixers have a leakage chamber between the oil chamber and the
gear unit which can be fitted with a leakage sensor. A leakage sensor detects fluid in
the leakage chamber resulting from a mechanical seal defect. The leakage sensor has
a separate cable for connection to an electrode relay. Tripping of the electrode relay
must result in the submersible mixer cutting out.
The electrode relay (K1) must meet the following requirements:
▪
Sensor circuit 10 to 30V ~
▪
Tripping current 0.5 to 3 mA
(equivalent to a tripping resistance of 3 to 60 k
Ω)
5.4.2 Connection to power supply
DANGER
Electrical connection work by unqualified personnel
Danger of death from electric shock!
▷
Always have the electrical connections installed by a trained electrician.
▷
Observe regulations IEC 30364 (DIN VDE 0100) and, for explosion-proof models,
IEC 60079 (DIN VDE 0164).
WARNING
Incorrect connection to the mains
Damage to the mains network, short circuit!
▷
Observe the technical specifications of the local energy supply companies.
CAUTION
Improper routing of power cable
Damage to the power cable!
▷
Never move the power cable at temperatures below - 25 °C.
▷
Never kink or crush the power cable.
▷
Never suspend the submersible mixer by its power cable.
5 Installation at Site
Amaprop®
21 of 60
Goodnal STP ST041 Biological Treatment (KSB Amaprop Installation and Operating Manual) Vendor Manual
Q-Pulse Id VM386
Active 29/10/2013
Page 21 of 60