Environmental chambers TC30 and TC40
page 73
3. Connect an inert gas line to the connection socket.
!
Do not set too high a value for the gas flow, as this could result in temperature
variations.
A good guideline value is 500 mbar.
Protective gas with condensation protection
If condensation protection is necessary, then also connect the protective gas to
the connectors provided for the purpose.
11.5.5
TC40 only: Power supply / Communication
!
Make sure that the control unit is switched off.
1. Plug the mains cable into the socket at the rear of the control unit.
2. Connect the control unit to the mains supply.
3. Connect the cable coming from the TC40 to the
HEATER
socket on the front of
the control unit.