4
OPERATION
46
UFC 400
www.krohne.com
12/2015 - 4002813702 - AD FF UFC 400 R02 en
SENSOR_CAL_
SENSOR_CAL_
SENSOR_CAL_
SENSOR_CAL_
WHO
WHO
WHO
WHO
Sensor Calibration
Who
R/W
The name of the person responsible for the last sensor
calibration.
-
SENSOR_SN
SENSOR_SN
SENSOR_SN
SENSOR_SN
Sensor Serial
Number
R
The sensor serial number.
-
SENSOR_TYPE
SENSOR_TYPE
SENSOR_TYPE
SENSOR_TYPE
Sensor Type
R
The type of sensor.
Ultrasonic (time of
travel)
Settings:
Ultrasonic (time of travel)
SIZE_ENTRY
SIZE_ENTRY
SIZE_ENTRY
SIZE_ENTRY
Size entry
R/W
Switches between a free value or a table value for the input of
the diameter (Fct. D4.4).
table
Settings:
value in [mm] / table [mm , inch]
SNR
SNR
SNR
SNR
SNR
R/W
SNR (Fct. D1.10)
Filtered
Settings:
Raw / Filtered
T_EXP_COEFF
T_EXP_COEFF
T_EXP_COEFF
T_EXP_COEFF
T expansion coeff.
R/W
Sets the temperature expansion coefficient (Fct. D2.9).
1.15e-5
°
C
Settings:
0.0
…
0.0001 1/
°
C
TIME_CONSTANT
TIME_CONSTANT
TIME_CONSTANT
TIME_CONSTANT
Time Constant
R/W
Time constant on the Front End for flow-related
measurement (volume, mass, flow velocity) (Fct. C1.3.3).
0.0 s
Settings:
0.0
…
1000.0 s
TRACKING
TRACKING
TRACKING
TRACKING
Tracking
R/W
Tracking (Fct. D1.9)
Normal
Settings:
Normal / Fast
TRANSDUCER_
TRANSDUCER_
TRANSDUCER_
TRANSDUCER_
DIRECTORY
DIRECTORY
DIRECTORY
DIRECTORY
Transducer
Directory Entry
A directory that specifies the number and starting indices of
the data collections in the transducer block.
0
TRANSDUCER_
TRANSDUCER_
TRANSDUCER_
TRANSDUCER_
FREQ
FREQ
FREQ
FREQ
Transducer
Frequency
R/W
Sets the frequency of the transducer (Fct. D1.1).
1 MHz
Settings:
1 MHz / 2 MHz / 500 kHz
TRANSDUCER_
TRANSDUCER_
TRANSDUCER_
TRANSDUCER_
TYPE
TYPE
TYPE
TYPE
Transducer Type
R
Identifies the transducer that follows.
Other
TRANSDUCER_
TRANSDUCER_
TRANSDUCER_
TRANSDUCER_
TYPE_VER
TYPE_VER
TYPE_VER
TYPE_VER
Transducer Type
Version
R
Identifies the version of the transducer that follows. Format is
XXYY where XX is the major spec revision and YY is the
manufacturer revision.
0
Trigger_Level
Trigger_Level
Trigger_Level
Trigger_Level
Trigger Level
R/W
Trigger level (Fct. D1.5)
30.0 %
Settings:
-99.0
…
99.0 %
Trigger_Margin
Trigger_Margin
Trigger_Margin
Trigger_Margin
Trigger Margin
R/W
Trigger margin (Fct. D1.6)
5.0 %
Settings:
0.0
…
50.0 %
WEIGHT_1
WEIGHT_1
WEIGHT_1
WEIGHT_1
Weight 1
R/W
Sets the weight of path 1 (Fct. D2.6).
1.0
Settings:
0.0
…
1.0
Parameter
Parameter
Parameter
Parameter
DD name
Access
Description and settings
Initial Value