FD_OFFSPEC_MASK
FD_OFFSPEC_MASK
FD_OFFSPEC_MASK
FD_OFFSPEC_MASK
Offspec Mask
R/W
This parameter allows the user to suppress any single or
multiple conditions that are active, in this category, from
being broadcast to the host through the alarm parameter.
(11) (12) (13) (14)
(20) (21) (22) (23)
Setting:
(00) Check / (01) Electronics : Operation Info /
(08) Electronics : Power Fail / (10) Configuration : Totaliser /
(11) Process : Signal Unreliable / (12) Process : Signal Lost /
(13) Process : Empty Pipe / (14) Electronics : IO Connection /
(15) Process : Current input / (16) M Process /
(17) M Configuration / (18) M Electronics / (19) M Sensor /
(20) S Process / (21) S Configuration / (22) S Electronics /
(23) S Sensor / (24) C Process / (25) C Configuration /
(26) C Electronics / (27) C Sensor / (28) F Process /
(29) F Configuration / (30) F Electronics / (31) F Sensor
FD_OFFSPEC_PRI
FD_OFFSPEC_PRI
FD_OFFSPEC_PRI
FD_OFFSPEC_PRI
Offspec Priority
R/W
This parameter allows the user to specify the priority of this
alarm category.
0
Setting:
0...15
FD_RECOMMEN_ACT
FD_RECOMMEN_ACT
FD_RECOMMEN_ACT
FD_RECOMMEN_ACT
R
This parameter is a device enumerated summarization of the
most severe condition or conditions detected. The DD help
should describe by enumerated action, what should be done to
alleviate the condition or conditions.
Uninitialized
Setting:
(0):Not Initialized / (1):No Action Required / (2):Remove
Simulate Jumper! / (1)(01)(00):Keys can be used 60 sec. after
removal of optical interface. / (1)(01)(01):Check configuration
settings of display line 2. / (1)(01)(02):Check configuration
settings of display line 1. / (1)(01)(06): Set MODE_BLK.TARGET
at RB to AUTO / (1)(01)(07): Check Fieldbus Terminal and
Connection / (1)(01)(11): / (1)(01)(12): / (1)(01)(13): /
(1)(01)(14): / (1)(01)(15): / (1)(01)(30):After starting up the
converter will respond. / (1)(01)(31):Zero Calibration will stop
and normal behavior will follow. / (1)(02)(20): Select another
measuement for meas. page 2 / (1)(02)(21): Select another
measuement for meas. page 1 / (1)(02)(28): Select another
measuement for IO B or activate diagnosis meas. / (1)(02)(29):
Select another measuement for IO B or activate diagnosis
meas. / (1)(02)(30): Select another measuement for IO A or
activate diagnosis meas. / (1)(02)(31): Select another
measuement for IO A or activate diagnosis meas. /
(1)(03)(00):reserved / (1)(04)(00):reserved /
(1)(05)(00):reserved / (1)(06)(00):reserved /
(1)(07)(00):reserved / (1)(08)(00):reserved /
(1)(09)(00):reserved / (1)(10)(00):reserved / (1)(11)(28):Try to
establish a reasonable flow profile or check the fluid /
(1)(11)(29):Check the acoustic properties of the medium for
this path / (1)(11)(30):Check the acoustic properties of the
medium for this path /
Parameter
Parameter
Parameter
Parameter
DD name
Access
Description and settings
Initial Value