22
UFM
3030
4.07.01
PULSE/UNIT
Pulse value for volume flow rate unit, see Function 3.05.09
4.07.02
PULSE/UNIT
Pulse value for heat power rate unit, see Function 3.05.10
4.08.00
PULS WIDTH
Pulse output pulse width incorrect
Ensure condition pulse width
≤
0.5 x pulse period time is met.
4.08.01 PULS
WIDTH
Pulse width for frequencies
≤
10 Hz, see Function 3.05.11
4.09.00 HART
Current
output
range for HART incorrect. If Hart is activated the
minimum possible current should be 4 mA.
Ensure condition CURR 0 pct
≥
4 mA is met.
4.09.01 CURR
RANGE
Range of current output, see Function 3.04.03
4.09.02
CURR 0 pct
Current value for 0% scale, see Function 3.04.04
4.10.00
INP/OUTP
The digital input (C) and current output (I) are not allowed to be
switched on simultaneously. If the Profibus option is activated only
one of the following input/output functions can be used: digital input
(C), current output (I), pulse output (P). The current output is
deactivated by setting the function of current output to off en setting
the range of current output to 0-20mA.
4.10.01
INP FUNCT
Function of digital input, see Function 3.06.01
4.10.02
CURR FUNCT Function of current output, see Function 3.04.01
4.10.03 CURR
RANGE
Range of current output, see Function 3.04.03
4.10.04
PULS FUNCT
Range of pulse output, see Function 3.05.01
4.13.00
EPROM
EPROM checksum error, reset device.