3.6
Connection diagram examples
Current output
Active
Ri <= 680 ohm
Passive
For supply:
U = 15…24Vdc, I >= 22mA
Pulse output
Active Passive
R1>= 470 ohm,
R2 = U*R1/(V+ - U)
For supply: U<= 32Vdc, <= 24Vac
Digital input
Analog input
Active
Passive
For supply: U = 15 –
30Vdc, I >= 1,5 mA
4 Start-up
•
Check that the flowmeter has been correctly installed.
•
With separate systems, check before initial start-up that the correct converter (UFC 030 F) is
used with the correct flow sensor (UFS 3000).
•
Order No., see instrument nameplates
•
Meter size (DN), Function 3.1.5
•
Primary constant GK, Function 3.1.6
•
Flow direction, Function 3.1.7
•
When powered, the signal converter operates in the measuring mode. TEST, NO ERROR and
IDENT NO. _ _ _ _ _ _ _ of the signal converter appear in succession on the display. This is
followed by display of the actual flow rate and/or the internal count on a continuous or alternating
basis (depending on setting, see Function 3.03 Display or Function 1.02 Display).
UFM
3030
11