Connecting the signal converter 

3.1 

3.2 

3.3 

Safety 

instructions 

 
This product is designed for use in accordance with EN IEC 61010-1 for Installation Category 2 and 
Pollution Degree 2. Hazardous voltages are present within this product during normal operation. The 
product is designed for Protection Class I and should never be operated without protective earthing. 
The product shall also never be operated with covers removed unless equivalent protection of the 
operator and its environment from accidental contact with hazardous internal voltages is provided. 
Always follow basic and local safety precautions when using this product to reduce risk of injury from 
electrical shock, spread of fire or other dangerous situations. 
 

Converter terminal box 

 

  The converter terminal box is accessible after removing the rear (blind) cover of the electronics 

section using the special wrench supplied with the flowmeter. 

  Do not damage the screw thread and the gasket, never allow dirt to accumulate, and make sure 

that the screw thread is well greased, using Teflon grease at all times. A damaged gasket must 
be replaced immediately! 

  Do not cross or loop the cables in the terminal box of the signal converter. Use separate cable 

entries for power supply and signal cables. 

  Special regulations apply to installation in hazardous areas (see installation instructions for 

hazardous areas). 
 

Power 

supply 

connection 

 
In case of connection to the mains supply voltage: 
 
Environmental conditions 

  The UFM3030 is designed to operate safe under the following conditions: 

  Suitable for indoor and outdoor use, the instrument is usable up to protection category  IP67 

(IEC 60529) 

  Use up to an altitude of 2000 m above see level 

  Suitable for an operation ambient temperature range  - 40 to +65°C 

  Suitable for an storage temperature range   -40 to + 80°C 

  Suitable for use in atmospheres with a relative humidity up to 80%  

  Over voltages up to category II on the main supply voltage ( IEC 60364-4-443) 

  Connected to protective earth conductor ( Protection Class I) 

  Rated pollution degree 2 

 
 
 

 

  This instrument is intended for permanent connection to the mains. It is required 

(for example for service) to mount an external switch or circuit breaker near the 
product for disconnection from the mains. It must be easily reachable by the 
operator and marked as the disconnecting device for this product.  
The switch or circuit breaker has to be suitable for the application and shall also be 
in accordance with to local (safety) requirements and of the building installation. 
(IEC 60947-1/-3). 

 

  The protective conductor clamp terminal size M5, press-fitted in the terminal 

compartment (near the main connection terminals), shall always to be connected  
to the protective earth conductor of the mains supply. Conductors up to 4 mm²  
(11 AWG) be connected to this terminal. The diameter of the conductors of the 
mains supply, including the protective earth conductor shall be in accordance with 

 

 

Summary of Contents for UFC 030

Page 1: ...Heat metering Communications technology Switches counters displays and recorders Engineering systems solutions DIN A4 7 10028 31 00 KROHNE 06 2005 7 30934 32 00 Short form installation and operating...

Page 2: ...nstall the product securely and stable Install and connect cabling proper to exclude damage or harmful situations If the product does not operate normally refer to the service instructions or refer to...

Page 3: ...liable in any way for possible consequences of such changes Product liability and warranty Responsibility for suitability and intended use of this ultrasonic flowmeter rests solely with the user Impro...

Page 4: ...nd your local KROHNE representative A simple operating check of the electronics after unpacking and before permanent installation is advisable to ascertain whether it has suffered damage during shipme...

Page 5: ...flowmeter by the signal converter housing or the terminal box Check the weight of the flowmeter as indicated on the type plate before handling the unit When handling the flowmeter avoid hard blows jol...

Page 6: ...entreline and any adjacent wall of at least 0 5 m 1 6 ft Special installation requirements To avoid measuring errors and malfunctioning of the flowmeter due to gas or air inclusions or an empty pipe p...

Page 7: ...with electric corrosion protection are generally insulated inside and outside so that the fluid has no conductive connection to ground The flowmeter must be insulated from the pipe Note the following...

Page 8: ...for hazardous areas Power supply connection In case of connection to the mains supply voltage Environmental conditions The UFM3030 is designed to operate safe under the following conditions Suitable f...

Page 9: ...DC 18 32 V dc AC 24 V AC 10 15 N N1 Neutral power supply PE Protective ground connection FE Functional ground connection Protective conductor clamp terminal Conductors up to 4 mm 11 AWG need to be con...

Page 10: ...ication Function can be set via menu option 3 4 0 and 3 6 0 Current output I I 22 mA Rload 680 ohm Umax 15Vdc Digital input C low 0 5 VDC high 15 32 VDC Will be switched off when current output activa...

Page 11: ...the correct converter UFC 030 F is used with the correct flow sensor UFS 3000 Order No see instrument nameplates Meter size DN Function 3 1 5 Primary constant GK Function 3 1 6 Flow direction Functio...

Page 12: ...rate Velocity of Sound VOS Total Totalizer forward flow Total Totalizer reverse flow Total Totalizer sum and 4 Compass field for error indication 5 Operating keys for programming the signal converter...

Page 13: ...e required setting or programming task Function block 2 Test contains all available test functions This block can be accessed to check proper functioning off all converter hard and software Function b...

Page 14: ...easuring mode Menu mode Data level Go to the parameter setting mode function 1 00 00 OPERATION If access CODE 1 is activated CODE 1 must be entered first Using function 3 07 02 access CODE 1 can be ac...

Page 15: ...IT Pulse value per volume flow unit see Function 3 05 09 1 03 03 PULSE UNIT Pulse value per energy unit see function 3 05 10 2 00 00 TEST Main menu 2 00 00 Test functions 2 01 00 DISPLAY Submenu 2 01...

Page 16: ...US Gal hr bbls hr bbls day free user configurable unit 3 01 02 ZERO VALUE Zero value see Function 1 01 02 FIXED factory zero setting MEASURED zero calibration possible see Function 3 01 03 3 01 03 ZER...

Page 17: ...08 through 3 02 11 CORR T P temperature correction via input 1 pressure correction via input 2 see also Function 3 02 08 through 3 02 11 HEAT heat measurement see function 3 02 12 BATCH batch volume s...

Page 18: ...nction 1 02 02 A Totalizer can be selected here for displaying Available options are TOTAL OFF FORWARD REVERSE BOTH SUM BOTH SUM NO DISPLAY 3 03 04 TOTAL VOL Unit for volume Totalizer X10 m US Gallon...

Page 19: ...02 11 F R IND forward reverse flow indication VOS velocity of sound range defined in Function 3 01 11 and 3 01 12 DIG OUTPUT digital output see Function 3 05 03 BATCH OUTP batch output indication see...

Page 20: ...h key only YES entry with key and code 1 factory set on 9 x key see Function 3 07 03 3 07 03 CODE 1 Code 1 Press any 9 keystroke combination and then press the same combination again Each keystroke is...

Page 21: ...t value for 100 scale see Function 3 04 05 4 03 00 LF CUTOFF Low flow cut off range incorrect If low flow cut off is set to on the value for CUTOFF OFF is compared with the value of CUTOFF ON on Ensur...

Page 22: ...on 3 04 03 4 09 02 CURR 0 pct Current value for 0 scale see Function 3 04 04 4 10 00 INP OUTP The digital input C and current output I are not allowed to be switched on simultaneously If the Profibus...

Page 23: ...he liquid used KROHNE regrets that they can not service your flowmeter unless accompanied with such a statement The following specimen statement is available on the KROHNE website as a word file Simpl...

Reviews: