11/2022 – 4009097302 - QS SMARTPAT PH R02 en 

 

 

Connection to SD 200 W 

 Field device 

 SD 200 W 

 Power supply / Control room 

 U+ 

  

 Sensor- 

 U- 

 
 
 
 
 

Connection to SHD 200 / SMARTMAC 400 

 

 

 

 

 

 

 SHD 200 / SMARTMAC 400, connection with or without internal resistor 

 Sensor with VP2-S cable 

 Relays 

 Loop 2 

(refer to manual for functionality)

 

 

HART connections for full sensor access

 

 

Control system without 250 Ohm resistor connected to internal resistor of SHD 200

 / SMARTMAC 400 

 Control system with 250 Ohm resistor connected to SHD 200 / SMARTMAC 400 without using internal resistor

 

SHD 200 / SMARTMAC 400 with internal resistor 

SHD 200 / SMARTMAC 400 without internal resistor 

Loop 2+ 

Loop 2+ 

Loop 2- 

Loop 2- 

FE 

Functional earth 

FE 

Functional earth 

Shield 

Shield 

A1 

 

A1 

 

B1 

Sensor- 

B1 

Sensor- 

A2 

HART+ 

A2 

HART+ 

B2 

HART- 

B2 

HART- 

Loop 1+ (HART

®

 resistor on board) 

 

 

Loop 1+ (external HART

®

 resistor required) 

Loop 1- 

Loop 1- 

Relay 1+ 

Relay 1+ 

Relay 1- 

Relay 1- 

Relay 2+ 

Relay 2+ 

Relay 2- 

Relay 2- 

Connection with OPTIBRIDGE and optional M8-VP cable 

 

 

 

 Primary master (PC / notebook with PACTware

TM

 FDT/DTM) 

 USB-connector 

 

M8 cable connector (female)

 

 M8 connector male 

 VP cable connector 

 Sensor connector 

 
Connection to SMARTMAC 200 W 

 
 
 

 

 

 
 
 
 

 SMARTMAC 200 W, connection with or without internal resistor 

 Sensor with VP2-S cable 

 Control system without 250 Ohm resistor 

connected to internal resistor of SMARTMAC 200 W

 

 Control system with 250 Ohm resistor connected to SMARTMAC 200 W without using internal resistor 

SMARTMAC 200 W with internal resistor 

SMARTMAC 200 W without internal resistor 

Shield 

Shield 

 

 

Sensor - 

Sensor - 

Loop - 

Loop - 

Loop + (250

Loop +  

 

4

5

6

7

J

K

FE

S

A1 B1

A2 B2

I

U2+

U2-

S

U1+

U1-

Sens-

UR1+ UR1-

UR2+

UR2-

J

K

FE

S

A1 B1

A2 B2 

C

U2+

U2-

S

U1+

U1-

Sens-

UR1+ UR1-

R2-

U

D

F

H

E

G

I

D

F

H

E

G

Sens+

Sens+

UR2+

Summary of Contents for SMARTPAT PH

Page 1: ...ntenance may only be performed by appropriately trained personnel Check the nameplate for correct operating conditions For use in hazardous areas special codes and regulations are applicable Instruments must not be connected to power supply before reading instructions described in the supplementary manual The responsibility as to the suitability intended use and corrosion resistance of the used ma...

Page 2: ...ART C Loop 1 HART resistor on board C D D Loop 1 external HART resistor required E Loop 1 E Loop 1 F Relay 1 F Relay 1 G Relay 1 G Relay 1 H Relay 2 H Relay 2 I Relay 2 I Relay 2 Connection with OPTIBRIDGE and optional M8 VP cable Primary master PC notebook with PACTwareTM FDT DTM USB connector M8 cable connector female M8 connector male VP cable connector Sensor connector Connection to SMARTMAC 2...

Page 3: ...d Buffer not recognised appear press Ok and enter pH value of buffer solution manually press Enter clean sensor with demineralised water dip sensor into second solution wait for stable value press Ok after 10 sec pH temperature Status buffer detection Finished and identified buffer value appear press Ok if necessary change pH buffer solution value manually press Enter if identified buffer 0 pH and...

Page 4: ...th the measuring medium The mounting position of the sensor should not deviate more than 75 from vertical position sensor tip pointing downwards Doing otherwise might cause internal air bubbles to float into the sensor tip This would interrupt the electrical contact between the inner buffer solution and the glass surface Calibration with PACTwareTM HART Handheld 475 FIELD COMMUNICATOR or SMARTMAC ...

Reviews: