11/2022 – 4009097302 - QS SMARTPAT PH R02 en 

 

 

3 Calibration 

Calibration with PACTware

TM

 (download free of charge on KROHNE.com) e.g. with OPTIBRIDGE

 

Manual hold 

select Menu mode 

Quick Setup

 or 

Setup

 / start 

calibration

 / activate 

manual hold

 / 

select 

yes

 / press 

Next

 

Temperature 
compensation 

set 

temperature compensation

 / select 

automatic

 or 

manual

 

(for 

manual

 enter 

temperature of pH buffer solution; default setting: 25°C (77°F) / press 

Next

 

Start calibration 
procedure 

dip sensor into first solution 

 

/ wait for stable value / press 

Next

 

/ after 

 

10 sec. 

pH-

temperature-

 and 

identified buffer values

 appear / press 

Next

 

/ if 

necessary, change pH buffer solution value manually / if 

buffer not recognised

 

appears press 

Next

 

and enter pH value of buffer solution manually / press 

Next

 / clean 

sensor with demineralised water 

 

dip sensor into second solution 

 

/ wait for st

able value / press 

Next

 

/ after 

10 sec. 

pH-

temperature-

 and 

identified buffer values

 appear / press 

Next

 

/ if necessary, 

change pH buffer solution value manually / if 

buffer not recognised

 appears press 

Next

 

and enter pH value of buffer solution manuall

y / press 

Next

 / clean sensor with 

demineralised water 

 

 

 

 

 

 

 

 

 

After successful 
calibration  

the following values appear: 

 

Buffer solution 1

Buffer solution 2

Offset [mV]

Slope [mV/pH]

 

press 

Next

 / set calibration date DD-MM-YYYY / set calibration time  

HH-MM-SS

 / press 

Next

 / 

Save values?

 / select 

yes

 / press 

Next 

/ select 

no

 to 

deactivate 

manual hold

 / press 

Next

 to leave calibration menu 

After unsuccessful 
calibration 

message 

Warning! Offset out of range!

 appears behind the value 

Offset

 and/or 

the message 

Warning! Slope out of range!

 behind the value 

Slope

. In case the 

value of the slope is not plausible, additionally the message 

No saving possible

 

appears. 

 

 

 

 

Calibration with HART Handheld 475 FIELD COMMUNICATOR

 

Manual hold 

select Menu mode 

Quick Setup

 or 

Setup

 / start 

calibration

 / activate 

manual 

hold

 / select 

yes

 / press 

Enter

 

Temperature 
compensation 

set 

temperature compensation

 / select 

automatic

 or 

manual

 

(for 

manual

 enter 

temperature of pH buffer solution; default setting: 25°C (77°F) / press 

Enter

 

Start calibration 
procedure 

dip sensor into first solution 

 

/ wait for stable value / press 

Ok

 

/ after 

10 sec. 

pH-

temperature-

Status buffer detection 

 Finished

 and 

identified buffer value

 

appear / press 

Ok

 

/ if necessary, change pH buffer solution value manually / press 

Enter

 

/  if 

identified buffer 

 

0 pH

 and 

Buffer not recognised

 appear press 

Ok

 

and enter pH value of buffer so

lution manually / press 

Enter

 / clean sensor with 

demineralised water 

 

dip sensor into second solution 

 

/ wait for stable value / press 

Ok

 

/ after 

10 sec. 

pH-

temperature-

Status buffer detection 

 Finished

 and 

identified buffer 

value

 appear / press 

Ok

 

/ if necessary, change pH buffer solution value manually / 

press 

Enter

 

/ if 

identified buffer 

 

0 pH

 and 

Buffer not recognised

 appear press 

Ok

 

and enter pH value of buffer solution manually / press 

Enter

 / clean sensor with 

demineralised water 

 

After successful 
calibration  

the following values appear: 

 

Buffer solution 1

Buffer solution 2

Offset [mV]

Slope [mV/pH]

 

Press 

Ok

 / set calibration date 

DD-MM-YYYY

 with key 

>

 / press 

Enter

 / press 

Ok

 / 

Save values?

 / select 

yes

 / press 

Enter

 / select “

no

” to deactivate “

manual hold

” / 

press 

Enter

 to leave calibration menu 

After unsuccessful 
calibration 

message 

Warning!

 appears behind the values 

Offset

 and/or 

Slope

If the message 

Attention! Slope out of range!

 appears, no data can be saved. 

 

 

 

Calibration with SMARTMAC 400

 

Manual hold 

select Menu mode 

PAT sensor

 / select 

Calibration

 / select 

Two point calibration 

All outputs on hold?

 / select 

Yes 

or

 No

 / press 

Enter

 

Temperature 
compensation 

Temperature comp. 

/ select 

Automatic

 or 

Manual

 

(for 

Manual

 enter temperature 

of pH buffer solution; default setting: 25°C (77°F)

) / press 

Enter

 

Start calibration 
procedure 

Buffer solution 1

 (e.g. 7.00 pH) 

/ dip sensor into buffer solution 

 

/ wait for stable 

value / press 

Enter

 / 

Set buffer solution 1

 (

if necessary, change pH value of used 

buffer solution manually

) / press 

Enter 

/ clean sensor with demineralised water 

 

Buffer solution 2 

(e.g. 4.00 pH) 

/ dip sensor into buffer solution 

 

/ wait for stable 

value / press 

Enter

 / 

Set buffer solution 2

 (

if necessary, change pH value of used 

buffer solution manually

) / press 

Enter

 / clean sensor with demineralised water 

 

After successful 
calibration  

the following

 is displayed:  

Calibration parameter

  

Slope: .….. [mV/pH] 
Offset: …… [mV] 
Calibration ok 

(if

 Slope in warning Range 

is displayed, the calibration can be 

stored, but it is recommended to 

check the buffer solutions and to 

repeat the 

procedure)

 

Press 

Enter

 / select 

yes

 / press 

Enter

 / 

Set calibration date

 

YYYY-MM-DD / 

press 

Enter /

 

Primary value .… pH

 and 

All outputs on hold?

 

select

 Yes 

or

 No / 

press

 Enter

 / 

Calibration successful / 

press

 Enter

 several times 

After unsuccessful 
calibration 

the following is displaye

d: 

Calibration parameter

   

Slope: ..… [mV/pH] 
Offset: ..… [mV] 
Slope out of range 

/

 

press 

Enter 

Primary value .… pH

 and 

All outputs on 

hold?

 / select 

Yes 

or

 No

 / press 

Enter 

Calibration aborted / 

press

 Enter

 several 

times

 

 

Summary of Contents for SMARTPAT PH

Page 1: ...ntenance may only be performed by appropriately trained personnel Check the nameplate for correct operating conditions For use in hazardous areas special codes and regulations are applicable Instruments must not be connected to power supply before reading instructions described in the supplementary manual The responsibility as to the suitability intended use and corrosion resistance of the used ma...

Page 2: ...ART C Loop 1 HART resistor on board C D D Loop 1 external HART resistor required E Loop 1 E Loop 1 F Relay 1 F Relay 1 G Relay 1 G Relay 1 H Relay 2 H Relay 2 I Relay 2 I Relay 2 Connection with OPTIBRIDGE and optional M8 VP cable Primary master PC notebook with PACTwareTM FDT DTM USB connector M8 cable connector female M8 connector male VP cable connector Sensor connector Connection to SMARTMAC 2...

Page 3: ...d Buffer not recognised appear press Ok and enter pH value of buffer solution manually press Enter clean sensor with demineralised water dip sensor into second solution wait for stable value press Ok after 10 sec pH temperature Status buffer detection Finished and identified buffer value appear press Ok if necessary change pH buffer solution value manually press Enter if identified buffer 0 pH and...

Page 4: ...th the measuring medium The mounting position of the sensor should not deviate more than 75 from vertical position sensor tip pointing downwards Doing otherwise might cause internal air bubbles to float into the sensor tip This would interrupt the electrical contact between the inner buffer solution and the glass surface Calibration with PACTwareTM HART Handheld 475 FIELD COMMUNICATOR or SMARTMAC ...

Reviews: