4
OPERATION
OPTIWAVE-M 7400 C
Status
type
Error message
Description
Corrective action
F
Configuration
Inconsistent NVRAM
Incorrect data in the parameter
memory.
De-energize and then energize the
device again.
If the message is shown again, tell
the supplier.
Incons. Converter Calib.
Incorrect calibration data in the
converter module.
Tell the supplier.
Incons. Sensor Calibration
Incorrect calibration data in the sensor
module.
NVRAMs Mismatch
The serial number of the display
screen does not agree with the serial
number of the electronics module.
Make sure that the display screen
agrees with the electronic module.
Go to Full Setup > Device >
Information and record the data in
menu items C7.1.5 Electronic Revision
and C7.1.6 Software Revision. If it is
necessary, tell the supplier.
Conv. NVRAM Layout Error
Incorrect data in the parameter
memory.
De-energize and then energize the
device again.
If the message is shown again, tell
the supplier.
Disp. NVRAM Layout Error
Incorrect data after an update of the
firmware.
CO Float Error
2
The device cannot use the
measurement data because the device
settings are incorrect.
You must make the device go back to
its factory settings. Enter Program
mode. Set the device to the "Expert"
access level. Go to Full Setup >
Device > Factory Default > Reset to
Factory Default? and set this menu
item to "Yes". Then go back to
Normal mode and set the "Save
Configuration?" page to "Yes". De-
energize and then energize the
device again.
If the message is shown again, tell
the supplier.
C
Electronics
FW Update
A firmware update of the converter
module continues.
Wait for the firmware update to
finish.