OPTIWAVE-M 7400 C
OPERATION
4
Menu
No.
Function
Function description
Selection list
Marine
Default
B1.1.5
Level Lin.
This sets the device to a given test level
value (linearized). This menu item is only
available if you set up a linearization table
in menu item C3.1 Edit Table. Make sure
that menu item C4.1.1 Current Out. 1 Var.
is set to "Lin. Level" before the device
simulates a level value. This procedure
sends an output signal that agrees with the
test reading. Output will change to the
selected value, independent of the
measurement data.
When you push [
^
] to confirm the value,
the display shows the question "Start
Simulation?" Push [
] or [
] to set the
display to "Yes". Push [
^
] again to start
the test. The device goes to back to
Normal mode after 1 hour.
Minimum access level to do the
procedure: Expert
min-max:
-5000.0...+5000.0 m /
-196.85
+03
...+196.85
+03
in /
-16404...+16404 ft
1
B1.1.6
Volume
This sets the device to a given test volume
value. This menu item is only available if
you set up a conversion (volume) table in
menu item C3.1 Edit Table. Make sure
that menu item C4.1.1 Current Out. 1 Var.
is set to "Volume" before the device
simulates a volume value. This procedure
sends an output signal that agrees with
the test reading. Output will change to
the selected value, independent of the
measurement data.
When you push [
^
] to confirm the value,
the display shows the question "Start
Simulation?" Push [
] or [
] to set the
display to "Yes". Push [
^
] again to start
the test. The device goes to back to
Normal mode after 1 hour.
Minimum access level to do the
procedure: Expert
min-max: 0
.... 1.00
+06
m
³
1
B1.1.7
Mass
This sets the device to a given test mass
value. This menu item is only available if
you set up a conversion (mass) table in
menu item C3.1 Edit Table. Make sure
that menu item C4.1.1 Current Out. 1 Var.
is set to "Mass" before the device
simulates a mass value. This procedure
sends an output signal that agrees with
the test reading. Output will change to
the selected value, independent of the
measurement data.
When you push [
^
] to confirm the value,
the display shows the question "Start
Simulation?" Push [
] or [
] to set the
display to "Yes". Push [
^
] again to start
the test. The device goes to back to
Normal mode after 1 hour.
Minimum access level to do the
procedure: Expert
min-max: 0
.... 10.000
+09
kg
1