27
6 Set up with the display and adjustment module
OPTISOUND 3020 C • Two-wire 4 … 20 mA/HART
30507-EN-190318
find a detailed description in the operating instructions manual of the
"
Display and adjustment module
".
HART mode
Standard
Address 0
The sensor measures the distance from the sensor to the product
surface. For indication of the real level, an allocation of the measured
distance to the percentage height must be carried out.
The actual level is then calculated on the basis of these entered
values. At the same time, the operating range of the sensor is limited
from maximum range to the requested range.
100%
0%
1 m
(39.37
")
5 m
(196.9
")
2
1
3
Fig. 27: Parameterisation example, Min./max. adjustment
1 Min. level = max. measuring distance
2 Max. level = min. measuring distance
3 Reference plane
The actual product level during this adjustment is not important,
because the min./max. adjustment is always carried out without
changing the product level. These settings can be made ahead of
time without the instrument having to be installed.
Proceed as follows:
1. Move from the measured value display to the main menu by
pushing
[OK]
.
▶
Basic adjustment
Display
Diagnostics
Service
Info
2. Select the menu item "
Basic adjustment
" with
[->]
and confirm
with
[OK]
. Now the menu item "
Min. adjustment
" is displayed.
Parameter adjustment
Basic adjustment - Min.
adjustment