2
INSTALLATION
12
OPTIFLEX 2200 C/F
www.krohne.com
03/2019 - 4005066003 - AD NEPSI OPTIFLEX2200 R03 en
Compact version
EPL Gb: Ex ia and Ex d ia devices
Temperature
class
Maximum ambient temperature
Max. flange
temperature
Ø2 mm / Ø0.08
¨
single cable probe
(HT version)
Ø2 mm / Ø0.08
¨
single cable probe
(other versions)
All other probe types
[
°
C]
[
°
F]
[
°
C]
[
°
F]
[
°
C]
[
°
F]
[
°
C]
[
°
F]
T6
+54
+129
+52
+125
+53
+127
+60
+140
+51
+124
+42
+107
+45
+113
+85
+185
T5
+69
+156
+67
+152
+68
+154
+75
+167
+66
+151
+57
+134
+60
+140
+100
+212
T4
+79
+174
+77
+170
+78
+176
+85
+185
+76
+169
+67
+152
+70
+158
+110
+230
+73
+163
+57
+134
+62
+143
+135
+275
T3
+71
+160
+51
+124
+57
+134
+150
+302
+68
+154
-
-
-
-
+180
1
+356
1
+65
+149
-
-
-
-
+200
1
+392
1
T2
+60
+140
-
-
-
-
+250
1
+482
1
+54
+129
-
-
-
-
+300
1
+572
1
1
Make sure that gasket temperature is in the specified limits. For more data, refer to the handbook.
Temperature
class
Minimum ambient temperature
Min. flange
temperature
Ø2 mm / Ø0.08
¨
single cable probe
(HT version)
Ø2 mm / Ø0.08
¨
single cable probe
(other versions)
All other probe types
[
°
C]
[
°
F]
[
°
C]
[
°
F]
[
°
C]
[
°
F]
[
°
C]
[
°
F]
T6...T2
-40
-40
-40
-40
-40
-40
-40
1
-40
1
-39
-38
-36
-32
-37
-34
-50
1
-58
1
1
Make sure that gasket temperature is in the specified limits. For more data, refer to the handbook.