INSTALLATION
2
7
OPTIFLEX 1100 C
www.krohne.com
01/2012 - 4001226201 - QS OPTIFLEX 1100 R01 en
2.2.2 How to assemble the segmented coaxial probe
•
Put the signal cable into the first segment of tube. Make sure that the signal cable goes
through the center hole in the spacer.
•
Attach the first segment of tube to the thread at the top of the signal cable. Tighten with a
17 mm open-end wrench.
•
Attach the lock screw to the top of the tube. Tighten with a 3 mm Allen wrench.
•
Do these steps again for each segment of tube, but do not attach the lock screw to the last
segment of tube.
Attach the tube segments
Figure 2-5: How to assemble the segmented coaxial probe: part 1
1
Top segment of tube
2
Bottom segment of tube
3
1 HC M6×5 lock screw
4
Bonded plastic spacer
CAUTION!
•
After you tighten the lock screw on each segment of tube, it will not be possible to
disassemble the coaxial probe.
•
If it is necessary to adjust the probe length, do not attach the lock screw to the last segment
of tube!