2
DEVICE DESCRIPTION
12
OPTIBAR 5060
www.krohne.com
04/2021 - 4005260702 - AD OPTIBAR 5060 Exta R02 en
Category 1/2/-D equipment
2.6.1 Ingress protection according to 60529
2.7 Electrical data
In order to guarantee the rated voltage U
m
and not to endanger the type of protection, the power
supply units used must meet the requirements for a SELF or PELV power supply unit.
2.7.1 OPTIBAR * 5060 VGK5/6/D*A/W/V R/H/7/S/T*******(*)**************
With electronics Z (4…20 mA), H (4…20 mA/HART
®
), A (4…20 mA/HART
®
with SIL
qualification)
With electronics H (4…20 mA/HART
®
), A (4…20 mA/HART
®
with SIL qualification) and
additional electronics Z (second current output)
Range
Permissible temperature range
Electronics housing
-40...+60°C / -40...+140°F
Max. surface temperature
Probe
Process tempe41 K
Table 2-9: Application for category 1/2/-D equipment
Protection class
Ingress protection
At probe, category 1D or 2D
IP68
At electronics housing, category 1D or 2D
IP68
OPTIBAR * 5060 VGK5/6/D*A/W/V R/H/7/S/T*******(*)Z/H/A/**************
Supply and signal circuit:
(Terminals 1[+], 2[-] in electronics
compartment, in the double chamber
housing version in terminal
compartment)
•
U
n
: 9.6...30 VDC
•
I
n
: 4...22 mA
•
U
m
: 30 VDC
Table 2-10: Electrical data for OPTIBAR * 5060 with electronics Z/H/A
OPTIBAR * 5060 VGK5/6/D*A/W/V R/H/7/S/T*******H/AZ*************
Supply and signal circuit I:
(Terminals 1[+], 2[-] in terminal
compartment)
•
U
n
: 9.6...30 VDC
•
I
n
: 4...22 mA
•
U
m
: 30 VDC
Supply and signal circuit II:
(Terminals 7[+], 8[-] in terminal
compartment)
•
U
n
: 9.6...30 VDC
•
I
n
: 4...22 mA
•
U
m
: 30 VDC
Table 2-11: Electrical data for OPTIBAR * 5060 with electronics Z/H and additional electronics Z