4/12
The cursor, flashing part of display, is shown below in bold type.
•
Change measuring range of current output and value for error messages (Fct. 1.05):
•
Change measuring range from 04-20 mA to 00-20 mA
•
Change value for error messages from 0 mA to 22 mA
Key
Display
Description
→
If ”YES“ set under Fct. 3.04 ENTRY CODE, key in the
9-keystroke CODE 1 now:
→ → → ↑ ↑ ↑ ↵ ↵ ↵
Fct. 1.00
OPERATION
→
Fct. 1.01
FULL SCALE
4x
↑
Fct. 1.05
CURRENT I
→
FUNCT. I
→ ↵
RANGE I
If ”REV. RANGE“ appears here,
press keys
→
and
↵
again.
→
04-20
mA
Old current range
2x
↑
00-20
mA
New current range
↵
I ERROR
→
0
mA
Old value for error messages
↑
22
mA
New value for error messages
↵
Fct. 1.05
CURRENT I
↵
Fct. 1.00
OPERATION
↵
STORE YES
↵
- - - - - - -
- - - - - / - - -
Measuring range with new data for the current output
Examples of setting the signal converter
4.7