
Photoc
hemical Hazar
d
(ISO 15004)
10
P10-1
The relative spectral output of the instrument
300
500
400
800
700
600
1100
1000
900
0.5
0.45
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
Wavelength (nm)
Image Capturing Light
Illumination Light
Energy( J) for image capturing light
Power( W) for illumination light
• The spectrally-weighted photochemical source radiance, both phakic L
B
and aphakic L
A
<Illumination Light>
<Image Capturing Light>
L
B
= 1.89 mW/(cm
2
• sr)
L
B
=66.8 mJ/(cm
2
• sr)
L
A
= 3.08 mW/(cm
2
• sr)
L
A
=98.2 mJ/(cm
2
• sr)
(informative)
Spectrally weighted photochemical radiances L
B
and L
A
give a measure of the potential that exists of a beam of
light to cause photochemical hazard to the retina. L
B
gives the measure for eyes in which the crystalline lens is in
place. L
A
gives this measure either for eyes in which the crystalline lens has been removed (aphakes) and has not
been replaced by a UV-blocking lens or for the eyes of very young children.
The value stated for this ophthalmic instrument gives a measure of hazard potential when the instrument is oper-
ated at maximum intensity and maximum aperture. Values of L
B
or L
A
over 80mW/(cm
2
• sr) are considered high for
beams which wholly fill a dilated pupil.
The retinal exposure dose for a photochemical hazard is a product of the radiance and the exposure time. For in-
stance, at a radiance level of 80mW/(cm
2
• sr), 3 min irradiation of the dilated (8mm diameter) pupil would cause
the retinal exposure dose level to attain the recommended exposure limit. If the value of radiance were reduced to
40mW/(cm
2
• sr), twice that time (i.e.6min) would be needed to reach the recommended limit. The recommended
exposure dose is based on calculations arising from the American Conference of Governmental Industrial Hygien-
ists (ACGIH) - Threshold Limit Values for Chemical Substances and Physical Agents (1995-1996 edition).
While no acute optical radiation hazards have been identified for ophthalmic instruments, it is recommended that
the intensity of light directed into the examined eye be limited to the minimum level which is necessary for diagno-
sis.Infants, aphakes and persons with diseased eyes will be at greater risk. The risk may also be increased if the
person being examined has had any exposure with the same instrument or any other ophthalmic instrument using
a visible light source during the previous 24 h. This will apply particularly if the eye has been exposed to retinal pho-
tography.
Should you have any question (e.g., ratio of the variable intensity to the maximum intensity), contact us or your
Kowa dealer.
Summary of Contents for VX-10a
Page 1: ...KOWA INSTRUCTION MANUAL FUNDUS CAMERA KOWA VX 10 EU...
Page 2: ......
Page 14: ...Introduction Introduction P0 12...
Page 63: ......