
2. Preparatory procedure
Pr
epar
a
tor
y
2
P2-5
Caution
Obligatory
Carefully adjust the height of the examined eye by moving the chin rest vertically while looking at
the patient’s face. There is a fear that too small face may be caught.
Warning
Prohibitory
Carefully operate the fundus camera so that the camera will not get in touch with the patient’s
face (eye, nose, so forth).
Caution
Prohibitory
Be careful for a patient to be unable to touch the instrument except for grips and a head band.
In operating the fundus camera, keep your fingers off the spaces or gaps between (1) the base
and power unit, (2) the main camera unit and horizontal arm, (3) tilt arm and tilt moving portion,
(4) the power unit and chin rest support, (5) the power unit and horizontal arm, and (6) the objec-
tive lens and the patient’s face. Otherwise, your fingers may be caught and injured. Instruct the
patient not to place his or her hand on the fundus camera.
Summary of Contents for VX-10a
Page 1: ...KOWA INSTRUCTION MANUAL FUNDUS CAMERA KOWA VX 10 EU...
Page 2: ......
Page 14: ...Introduction Introduction P0 12...
Page 63: ......