30
No response from
instrument
Communication with
the instrument failed.
• The power of the
instrument is off.
• COM port setting
is incorrect.
• Communication
settings are incorrect.
Turn on the power of the
instrument.
Check and correct
the COM port and
communication settings.
15
Hardware initialization
at power on (of lens,
filter and trap positions)
is in progress.
After turning the power on,
wait for ten and several
seconds until the lens, trap
and UV filter positions are
initialized.
-
Time for Periodic
Calibration. Please
contact the closest
Service Center.
It is time to perform
periodic calibration.
Contact a KONICA
MINOLTA-authorized
service facility and request
periodic calibration.
-
Not Calibration data.
Set Calibration data.
Are you sure?
The calibration data for
while calibration is not found.
Load the calibration data
into the instrument.
20
Check UV Settings.
UV Settings has been
set to "100% Full".
Set the UV Settings to
"UV adjustment" .
Refer to the
SpectraMagic™
NX (optional)
instruction manual.
UV adjustment not
possible.Change
Standard value or
tolerance settings,
or use a different
Fluorescent Standard.
Current settings did
not allow adjustment
of the filter position.
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard;
then retry "UV adjustment".
34
UV adjustment not
possible. UV is set
to XX and cannot be
adjusted further in
the required direction.
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard.
Current settings did
not allow adjustment
of the filter position.
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard;
then retry "UV
adjustment".
34