27
WARNING
Do not disassemble or modify the CM-3700A or AC adapter.
Doing so may cause a fire or electric shock.
The CM-3700A should not be operated if it is damaged, or if smoke or strange odors occur. Doing
so may result in a fire. In such situations, turn the power OFF immediately, disconnect the AC
adapter from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
CAUTION
When cleaning, disconnect the power plug. Connecting may cause electric shock.
Be careful around openings in the CM-3700A. Failure to do so
may result in fingers being trapped causing injury.
Zero Calibration Box and White Calibration Plate
Wipe gently with a dry soft cloth. If dirt is difficult to remove, dampen a cloth with commercially available lens
cleaning liquid and wipe. Then wipe off the liquid with a cloth dampened with water, and leave it to dry.
{
When cleaning, take care not to scratch the zero calibration box or white calibration plate.
Target Mask
Use a blower to remove dirt and dust from the target masks.
{
Do not touch the black-coated surface of the target masks with fingers or wipe it with a cloth. If the target masks
get so dirty that dirt cannot be removed using a blower, contact the nearest Konica Minolta-authorized service
facility.
Inside Integrating Sphere
1.
Remove the light trap cover. (See P.26)
Remove the light trap cover. (See P.26)
{
Be sure to set the SCI/SCE setting to SCI before
removing the light trap.
{
Store the light trap and screw carefully.
2.
Make sure that there is nothing placed
against the illumination window.
3.
Block the receiving window so that no dust
or dirt enters.
4.
Set the specular component to SCE using the
software.
5.
Open the Measurement port baseplate and
remove dust and dirt using a blower.
{
Do not touch the white-coated inner surface of the
integrating sphere, wipe it with a cloth or place any
object against it. If the white-coated surface gets
so dirty that dirt cannot be removed using a blower,
contact the nearest Konica Minolta-authorized
service facility.
Cleaning the CM-3700A and Accessories
Receiving window
Illumination window