26
Error Messege
Problem / possible cause
Action
Refer to page
Finder port is open.
The viewfinder cover stays
open during measurement.
Close the viewfinder cover
before measurement.
The measured object does
not exhibit fluorescence.
UV adjustment is not allowed
due to the UV specimen not
containing fluorescence.
Perform UV adjustment
using a specimen that
contains fluorescence.
—
The fluorescent calibration
coefficients could not
be determined.
The fluorescent
calibration coefficient
cannot be determined.
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard;
then retry "UV adjustment".
—
No response from instrument Communication with
the instrument failed.
• The power of the
instrument is off.
• COM port setting
is incorrect.
• Communication
settings are incorrect.
Turn on the power of
the instrument.
Check and correct
the COM port and
communication settings.
Detected instrument
disagrees with connected
one.\nConnect again.
The instrument is incorrect. Connect CM-3600A.
—
File is different to
current status.
The measurement
conditions in the fluorescent
calibration coefficient file
(.krd) designated to be read
are different from the actual
measurement conditions.
Adapt the actual
measurement conditions
to the measurement
conditions in the .krd
file or select a file that
corresponds with the actual
measurement conditions.
SpectraMagic™
NX (optional)
instruction
manual.
UV adjustment not
acceptable.
UV adjustment with a
Ganz&Griesser option has
not yet included all the
required measurements.
Measure the fluorescent
sample four times when
selecting "Ganz&Griesser 4"
or five times when selecting
"Ganz&Griesser 5".
SpectraMagic™
NX (optional)
instruction
manual.