29
29
Installation/Connection
Turning the Power ON ( | )/OFF (
●
)
1. Turning the Power ON ( |
)/OFF (
●
)
Before setting the POWER switch to ON ( | ), prepare the following.
1
.
Connect a measuring probe to the probe connector [P1]. (Page 26)
•
To synchronize measurement with the ...
1
Input the vertical synchronizing signal that is used for the display.
display’s vertical synchronizing signal
(Page 28)
(EXT is selected as the SYNC mode)
•
To perform measurement .......................
1
Install the 4-Probe Expansion Board CA-B04 (option) in the
with two or more measuring probes
instrument. (Page 27)
2
Connect necessary number of probes to the probe connec-
tors [P2] to [P5]. (Pages 26 and 27)
•
To communicate with the PC .................
1
Connect the instrument’s RS-232C connector to the PC. (Page 84)
via RS-232C
•
To communicate with the PC via USB ...
1
Connect the instrument’s USB connector to the PC. (Page 86)
2
. Connect the AC power cord to an AC outlet. (Page 28)
[Turning the Power ON ( | )]
Set the POWER switch to ON ( | ).
If the instrument is connected to external equipment,
set the instrument’s POWER switch to ON ( | ) first,
then turn ON ( | ) the power to the external equipment.
[Turning the Power OFF (
●
)]
If the instrument is connected to external equipment, turn OFF (
●
) the power to the external equipment
first, then set the instrument’s POWER switch to OFF (
●
).
<Error Messages in LCD Display Section>
… For other error messages, refer to page 107.
●
“SET MAIN PROBE” (After the POWER switch is set to ON ( | ))
•
Cause 1 : The measuring probe is not connected to the probe con-
nector [P1] properly.
•
Action 1: Set the POWER switch to OFF (
●
), then connect the measuring probe to the probe connector
[P1] properly. (Before connecting/disconnecting the measuring probe, make sure that the
POWER switch is set to OFF (
●
).)
●
“PROBE ERROR”
•
Cause 1 : A measuring probe was connected or disconnected while
the POWER switch was ON ( | ).
•
Action 1: Set the POWER switch to OFF (
●
) first, connect necessary measuring probes, then set the
POWER switch to ON ( | ). (Before connecting/disconnecting the measuring probe, make sure
that the POWER switch is set to OFF (
●
).)
PROBE
[P1]
NO.XXXXXXXX
DARKEN PROBE
PUSH 0-CAL KEY
CA-200 MODE
Probe serial no.
SET MAIN PROBE
Either “CA-200
MODE” or “CA-100
MODE” is displayed
(see page 13).
PROBE ERROR
Summary of Contents for CA-100PLUS
Page 24: ...22...
Page 46: ...44...
Page 76: ...74...
Page 100: ...98...
Page 108: ...106 Dimension Diagram Main Body Measuring Probe 60 ISO 5mm screw depth 6mm...
Page 123: ...121 Explanation Section...
Page 124: ......
Page 125: ...9222 1891 13 2002 KONICA MINOLTA SENSING INC AEIDPK 10 Printed in Japan...